Wall Shear-Rate Estimation Within the 50cc Penn State Artificial Heart Using Particle Image Velocimetry
Authors:
- 1 August 2004
- journal article
- research article
- Published by ASME International in Journal of Biomechanical Engineering
- Vol. 126 (4) , 430-437
Abstract
Particle image velocimetry (PIV) has been gaining acceptance as a routine tool to evaluate the flow fields associated with fluid mechanical devices. We have developed algorithms to investigate the wall shear-rates within the 50cc Penn State artificial heart using low magnification, conventional particle image velocimetry (PIV). Wall shear has been implicated in clot formation, a major post-implant problem with artificial hearts. To address the issues of wall scattering and incomplete measurement volumes, associated with near wall measurements, we have introduced a zero masking and a fluid centroid shifting technique. Simulations using different velocity fields were conducted with the techniques to assess their viability. Subsequently, the techniques were applied to the experimental data collected. The results indicate that the size of the interrogation region should be chosen to be as small as possible to maximize resolution while large enough to ensure an adequate number of particles per region. In the current study, a 16×16 interrogation window performed well with good spatial resolution and particle density for the estimation of wall shear rate. The techniques developed with PIV allow wall shear-rate estimates to be obtained from a large number of sites at one time. Because a planar image of a flow field can be determined relatively rapidly, PIV may prove useful in any preliminary design procedure.Keywords
This publication has 30 references indexed in Scilit:
- Near-wall physics of a shear-driven three-dimensional turbulent boundary layer with varying crossflowJournal of Fluid Mechanics, 2003
- Five techniques for increasing the speed and accuracy of PIV interrogationMeasurement Science and Technology, 2001
- Antithrombotic Efficacy of Inactivated Active Site Recombinant Factor VIIa Is Shear Dependent in Human BloodArteriosclerosis, Thrombosis, and Vascular Biology, 1997
- LDA Measurements of Mean Velocity and Reynolds Stress Fields Within an Artificial Heart VentricleJournal of Biomechanical Engineering, 1994
- Particle-Imaging Techniques for Experimental Fluid MechanicsAnnual Review of Fluid Mechanics, 1991
- Hot-Film Wall Shear Probe Measurements Inside a Ventricular Assist DeviceJournal of Biomechanical Engineering, 1988
- Pulsed Ultrasonic Doppler Velocity Measurements Inside a Left Ventricular Assist DeviceJournal of Biomechanical Engineering, 1986
- Clinical Use of the Total Artificial HeartNew England Journal of Medicine, 1984
- LASER DOPPLER ANEMOMETER STUDIES IN UNSTEADY VENTRICULAR FLOWSASAIO Journal, 1979
- ARTIFICIAL HEART EVALUATION USING FLOW VISUALIZATION TECHNIQUESASAIO Journal, 1972