Fast detection and data compensation for electrodes disconnection in long-term monitoring of dynamic brain electrical impedance tomography
Open Access
- 7 January 2017
- journal article
- research article
- Published by Springer Nature in BioMedical Engineering OnLine
- Vol. 16 (1) , 7
Abstract
Electrode disconnection is a common occurrence during long-term monitoring of brain electrical impedance tomography (EIT) in clinical settings. The data acquisition system suffers remarkable data loss which results in image reconstruction failure. The aim of this study was to: (1) detect disconnected electrodes and (2) account for invalid data. Weighted correlation coefficient for each electrode was calculated based on the measurement differences between well-connected and disconnected electrodes. Disconnected electrodes were identified by filtering out abnormal coefficients with discrete wavelet transforms. Further, previously valid measurements were utilized to establish grey model. The invalid frames after electrode disconnection were substituted with the data estimated by grey model. The proposed approach was evaluated on resistor phantom and with eight patients in clinical settings. The proposed method was able to detect 1 or 2 disconnected electrodes with an accuracy of 100%; to detect 3 and 4 disconnected electrodes with accuracy of 92 and 84% respectively. The time cost of electrode detection was within 0.018 s. Further, the proposed method was capable to compensate at least 60 subsequent frames of data and restore the normal image reconstruction within 0.4 s and with a mean relative error smaller than 0.01%. In this paper, we proposed a two-step approach to detect multiple disconnected electrodes and to compensate the invalid frames of data after disconnection. Our method is capable of detecting more disconnected electrodes with higher accuracy compared to methods proposed in previous studies. Further, our method provides estimations during the faulty measurement period until the medical staff reconnects the electrodes. This work would improve the clinical practicability of dynamic brain EIT and contribute to its further promotion.Keywords
This publication has 44 references indexed in Scilit:
- In Vivo Imaging of Twist Drill Drainage for Subdural Hematoma: A Clinical Feasibility Study on Electrical Impedance Tomography for Measuring Intracranial Bleeding in HumansPLOS ONE, 2013
- An analytic solution to the homogeneous EIT problem on the 2D disk and its application to estimation of electrode contact impedancesPhysiological Measurement, 2011
- The impact of electrode area, contact impedance and boundary shape on EIT imagesPhysiological Measurement, 2011
- An optimized strategy for real-time hemorrhage monitoring with electrical impedance tomographyPhysiological Measurement, 2011
- Real-Time Imaging and Detection of Intracranial Haemorrhage by Electrical Impedance Tomography in a Piglet ModelJournal of International Medical Research, 2010
- Real-time imaging of subarachnoid hemorrhage in piglets with electrical impedance tomographyPhysiological Measurement, 2010
- Pseudo-polar drive patterns for brain electrical impedance tomographyPhysiological Measurement, 2006
- A Practical Approach to Electrode-Skin Impedance Unbalance MeasurementIEEE Transactions on Biomedical Engineering, 2006
- Automatic detection of detached and erroneous electrodes in electrical impedance tomographyPhysiological Measurement, 2005
- A theory for multiresolution signal decomposition: the wavelet representationIEEE Transactions on Pattern Analysis and Machine Intelligence, 1989