We developed a compact synchrotron X‐ray topography system equipped with a demountable Si channel‐cut crystal monochromator with an energy range of 5–20 keV at beamline 09A in SAGA Light Source (SAGA‐LS) to enable switching between a white X‐ray beam and a monochromatic X‐ray beam. The position of incidence of the X‐ray beam on the sample is kept constant because the beam‐path difference, which is caused by inserting the monochromator and by tuning the X‐ray energy, is adjusted automatically. Two possible applications of this system are proposed: (1) the observation of dislocations in a thick Si rod by white X‐ray topography, and (2) a comparative study of X‐ray images of defects and strain in a 4H‐SiC wafer captured using white and monochromatic X‐ray beams.Keywords
This publication has 13 references indexed in Scilit:
- Current Status of the SAGA Light Source CommissioningAIP Conference Proceedings, 2007
- Design of Beamline BL9 at Saga Light SourceAIP Conference Proceedings, 2007
- Determination of the three-dimensional structure of dislocations in silicon by synchrotron white X-ray topography combined with a topo-tomographic techniqueJournal of Synchrotron Radiation, 2004
- SPECTRA: a synchrotron radiation calculation codeJournal of Synchrotron Radiation, 2001
- Influence of impurities on protein crystal perfectionJournal of Crystal Growth, 2001
- Back-reflection topographic study of mixed cells in LEC-grown GaAs 0.2 at.% InJournal of Physics D: Applied Physics, 1993
- Synchrotron X-radiation topographyMaterials Science Reports, 1992
- X-ray topography with synchrotron radiationNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1987
- A high-speed X-ray topography camera for use with synchrotron radiation at the photon factoryNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1984
- X-Ray Topographic Images of Oxidation-Induced Frank Sessile Dislocation Loops in Thin SiliconJapanese Journal of Applied Physics, 1981