Comparison of Dynamic Contour Tonometry and Goldmann Applanation Tonometry
Open Access
- 1 January 2009
- journal article
- Published by Korean Ophthalmological Society in Journal of the Korean Ophthalmological Society
- Vol. 50 (2) , 242-246
Abstract
AIM:To evaluate and compare intraocular pressure value measured by the dynamic contour tonometer and Goldmann applanation tonometer after laser in situ keratomileusis(LASIK) operation.METHODS:There were 34 myopic patients(68 eyes) who received LASIK.Intraocular pressures were measured in the preoperation and postoperative 3 months using Pascal dynamic contour tonometer(PDCT) and Goldmann applanation tonometer(GAT).Values measured by two methods were compared.Multiple linear correlation analysis was used to find the correlation between GAT,PDCT measurement,corneal curvature and central corneal thickness(CCT).RESULTS:After LASIK operation GAT measurement value was lower than that before treatment,and postoperative and preoperative difference in PDCT value was not statistically significant.The curvature of the cornea,CCT and GAT readings showed a linear correlation,but PDCT reading unrelated.CONCLUSION:GAT measurement of the intraocular pressure is lower than the actual value.PDCT intraocular pressure measurement is not affected by the corneal curvature and CCT.Keywords
This publication has 19 references indexed in Scilit:
- Effect of Central Corneal Thickness on Dynamic Contour Tonometry and Goldmann Applanation Tonometry in Primary Open-angle GlaucomaArchives of Ophthalmology (1950), 2007
- Effects of Corneal Thickness, Corneal Curvature, and Intraocular Pressure Level on Goldmann Applanation Tonometry and Dynamic Contour TonometryOphthalmology, 2007
- Effect of Corneal Thickness on Dynamic Contour, Rebound, and Goldmann TonometryOphthalmology, 2006
- Comparison of Dynamic Contour Tonometry with Goldmann Applanation TonometryInvestigative Ophthalmology & Visual Science, 2004
- Dynamic Contour Tonometry for Post-LASIK Intraocular Pressure MeasurementsKlinische Monatsblätter für Augenheilkunde, 2004
- Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIKJournal of Cataract & Refractive Surgery, 2004
- Central corneal thickness measurements in patients with normal tension glaucoma, primary open angle glaucoma, pseudoexfoliation glaucoma, or ocular hypertensionBritish Journal of Ophthalmology, 2001
- Human Corneal Thickness and Its Impact on Intraocular Pressure MeasuresSurvey of Ophthalmology, 2000
- A new tonometer—the pressure phosphene tonometerOphthalmology, 1998
- APPLANATION TONOMETRY AND CENTRAL CORNEAL THICKNESSActa Ophthalmologica, 1975