Performance of a new pulse contour method for continuous cardiac output monitoring: validation in critically ill patients
Open Access
- 26 April 2013
- journal article
- clinical trial
- Published by Elsevier in British Journal of Anaesthesia
- Vol. 111 (4) , 573-579
Abstract
A new calibrated pulse wave analysis method (VolumeView™/EV1000™, Edwards Lifesciences, Irvine, CA, USA) has been developed to continuously monitor cardiac output (CO). The aim of this study was to compare the performance of the VolumeView method, and of the PiCCO2™ pulse contour method (Pulsion Medical Systems, Munich, Germany), with reference transpulmonary thermodilution (TPTD) CO measurements. This was a prospective, multicentre observational study performed in the surgical and interdisciplinary intensive care units of four tertiary hospitals. Seventy-two critically ill patients were monitored with a central venous catheter, and a thermistor-tipped femoral arterial VolumeView™ catheter connected to the EV1000™ monitor. After initial calibration by TPTD CO was continuously assessed using the VolumeView-CCO software (CCOVolumeView) during a 72 h period. TPTD was performed in order to obtain reference CO values (COREF). TPTD and arterial wave signals were transmitted to a PiCCO2™ monitor in order to obtain CCOPiCCO values. CCOVolumeView and CCOPiCCO were recorded over a 5 min interval before assessment of COTPTD. Bland–Altman analysis, %errors, and concordance (trend analysis) were calculated. A total of 338 matched sets of data were available for comparison. Bias for CCOVolumeView−COREF was −0.07 litre min−1 and for CCOPiCCO–COREF +0.03 litre min−1. Corresponding limits of agreement were 2.00 and 2.48 litre min−1 (Perrors 29 and 37%, respectively. Trending capabilities were comparable for both techniques. The performance of the new VolumeView™-CCO method is as reliable as the PiCCO2™-CCO pulse wave analysis in critically ill patients. However, an improved precision was observed with the VolumeView™ technique. NCT01405040.Keywords
This publication has 40 references indexed in Scilit:
- Clinical validation of a new thermodilution system for the assessment of cardiac output and volumetric parametersCritical Care, 2012
- Lack of Effectiveness of the Pulmonary Artery Catheter in Cardiac SurgeryAnesthesia & Analgesia, 2011
- So we use less pulmonary artery catheters—But why?*Critical Care Medicine, 2011
- Pulmonary artery catheters: Evolving rates and reasons for use*Critical Care Medicine, 2011
- Cardiac output monitoring: an integrative perspectiveCritical Care, 2011
- Validation of a new transpulmonary thermodilution system to assess global end-diastolic volume and extravascular lung waterCritical Care, 2010
- Cardiac Output Monitoring Using Indicator-Dilution Techniques: Basics, Limits, and PerspectivesAnesthesia & Analgesia, 2010
- Minimally invasive haemodynamic monitoringEuropean Journal of Anaesthesiology, 2009
- Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomised controlled trialThe Lancet, 2005
- Functional hemodynamic monitoringCritical Care, 2005