3D X-Ray Source Deblurring in High Cone-Angle Micro-CT
- 16 July 2015
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 62 (5) , 2075-2084
Abstract
High geometric magnification X-ray micro-computed tomography (μCT) is used to study many high-resolution features in insects, cellular, bones, composite and mineral materials. The resolution of lab-based μCT in a fine-focus geometry is limited by blurring that occurs below the spatial coherence length of the illuminating radiation: resolution can be no smaller than the size of the X-ray source spot. In cases where the source spot size cannot be reduced (e.g. due to signal-to-noise, time or cost considerations) there is a need to model and correct for this blurring. In ANU CT-lab, we use a high cone angle and high geometric magnification with transmission x-ray source spot size up to three voxels, this creates blurring in the projection. This work takes a simulation approach mimicking such source spot size, and compares systems with horizontal cone-angles (often referred to as the fan angle) of 0.06, 14.36 and 60 degrees. We aim to eliminate this blurring in the reconstruction process. Furthermore, in a high cone-angle geometry, using a reconstruction method that only deconvolves each projection image leads to non-uniform resolution in the reconstruction volume. Alternatively, iterative methods that fully model the non-point source and avoid such artefacts are computationally expensive. We propose a hybrid method that corrects the effect of the non-point source by better modelling the physics rather than just deconvolving each projection image, therefore obtains results closer to the iterative full modelling method, and while being computationally much cheaper.Keywords
This publication has 17 references indexed in Scilit:
- Considerations for high-magnification high-cone-angle helical micro-CT2012
- Propagation beam consideration for 3D THz computed tomographyOptics Express, 2012
- Image reconstruction from a small number of projectionsInverse Problems, 2008
- Correction for Resolution Nonuniformities Caused by Anode Angulation in Computed TomographyIEEE Transactions on Medical Imaging, 2008
- Coherent X-Ray Optics2006
- The Mathematics of Computerized Tomography2001
- Optical Coherence and Quantum Optics1995
- A regularized iterative image restoration algorithmIEEE Transactions on Signal Processing, 1991
- Practical cone-beam algorithmJournal of the Optical Society of America A, 1984
- Function minimization by conjugate gradientsThe Computer Journal, 1964