Flexible, Conductive Fabric‐Backed, Microneedle Electrodes for Electrophysiological Monitoring
Open Access
- 8 December 2023
- journal article
- research article
- Published by Wiley in Advanced Materials Technologies
- Vol. 9 (3)
Abstract
Microneedle‐based electrodes have attracted significant attention due to their potential applications in high‐quality, long‐term electrophysiological monitoring. In this study, micromoulding technology is used to develop a flexible, conductive fabric microneedle (CF–MN) electrode by bonding a conductive textile to polymeric microneedle arrays. The resulting electrode conforms to the curved surfaces of the body, and the use of a conductive textile backing reduces metallization time and costs by 50% over double‐sided dry electrodes. The electrode‐skin impedance of the CF–MN electrodes is significantly lower than that of gel electrodes over a wide frequency range, and tests on healthy volunteers, both at rest and under a range of ambulatory conditions, showed that the electrodes are capable of acquiring electrocardiography (ECG) and electromyography (EMG) signals that have a comparable kurtosis and signal‐to‐noise ratio to those recorded using conventional electrodes, but without the need for skin preparation or the use of a wet gel. The biomechanical properties of the electrode are also characterized. This is the first demonstration of CF–MN electrodes that can ultimately be integrated into clothing, which in turn can facilitate personalized cardiac health management by enabling continuous, long‐term ECG/EMG monitoring during daily activities.Keywords
Funding Information
- Science Foundation Ireland (SFI/12/RC/2289‐P2)
- European Regional Development Fund
This publication has 49 references indexed in Scilit:
- Development of a Wearable Mobile Electrocardiogram Monitoring System by Using Novel Dry Foam ElectrodesIEEE Systems Journal, 2013
- The development and characteristics of novel microneedle arrays fabricated from hyaluronic acid, and their application in the transdermal delivery of insulinJournal of Controlled Release, 2012
- Design, Fabrication and Experimental Validation of a Novel Dry-Contact Sensor for Measuring Electroencephalography Signals without Skin PreparationSensors, 2011
- Wafer-scale fabrication of penetrating neural microelectrode arraysBiomedical Microdevices, 2010
- Silicon microneedle formation using modified mask designs based on convex corner undercutJournal of Micromechanics and Microengineering, 2006
- Application of independent component analysis in removing artefacts from the electrocardiogramNeural Computing and Applications, 2005
- The basics of electromyographyJournal of Neurology, Neurosurgery & Psychiatry, 2005
- Insertion of microneedles into skin: measurement and prediction of insertion force and needle fracture forceJournal of Biomechanics, 2004
- Temporal changes in electrode impedance while recording the electrocardiogram with “Dry” electrodesAnnals of Biomedical Engineering, 1973
- Recommendations for Standardization of Leads and of Specifications for Instruments in Electrocardiography and VectorcardiographyCirculation, 1967