Previously, investigators have measured endogenous digoxin-like immunoreactive substances (DLIS) in pediatric patients. Digoxin-like immunoreactive substances may cross-react with digoxin assays to produce false-positive digoxin concentrations; hence, the validity of digoxin concentrations in pediatric patients is questionable. The authors compared the presence and magnitude of apparent DLIS using the microparticle enzyme immunoassay (MEIA) AxSYM Digoxin II and the fluorescence polarization immunoassay (FPIA) TDx Digoxin II, in the serum of 80 pediatric patients who were hospitalized with normal serum creatinine but had not been administered digoxin. Patients ranged in age from 1 day to 16 years (mean age, 4.96± 5.17 years). Serum creatinine and total bilirubin were 0.5 ± 0.18 mg/dl and 1.3 ± 0.17 mg/dl, respectively. Forty-eight percent of MEIA samples and 79% of FPIA samples had measurable DLIS values. Digoxin-like immunoreactive substance concentrations for the MEIA (0.07 ± 0.09 ng/ml) and FPIA assays (0.1 ± 0.1 ng/ml) were statistically different(p = 0.01); however, no sample had a DLIS value >0.38 ng/ml. A poor correlation was noted between patient age, serum creatinine, total bilirubin, and DLIS concentration. The MEIA and FPIA assays effectively minimized DLIS cross-reactivity making both technologies clinically acceptable for serum digoxin measurement in pediatric patients with normal serum creatinine and total bilirubin.Keywords
This publication has 30 references indexed in Scilit:
- Endogenous Digoxin-like Immunoreactive Factor Is Elevated in Advanced Chronic Respiratory FailureChest, 1992
- Interference of Digoxin-Like Immunoreactive Substances with TDx Digoxin II Assay in Different PatientsTherapeutic Drug Monitoring, 1991
- The influence of gestational age and preeclampsia on the presence and magnitude of serum endogenous digoxin-like immunoreactive substance(s)American Journal of Obstetrics and Gynecology, 1988
- Endogenous digoxin-like immunoreactivity in congestive heart failure.BMJ, 1987
- Effect of age and serum creatinine on endogenous digoxin-like substances in infants and childrenThe Journal of Pediatrics, 1987
- Digoxin- and Digitoxin-Like Immunoreactive Substances in Amniotic Fluid, Cord Blood, and Serum of NeonatesPediatric Research, 1986
- Endogenous Digoxin-Immunoreactive Substance in Human Pregnancies*The Journal of Clinical Endocrinology & Metabolism, 1984
- Endogenous substance in newborn infants causingfalse positive digoxin measurementsThe Journal of Pediatrics, 1983
- A circulating inhibitor of (Na+ + K+) ATPase associated with essential hypertensionNature, 1982