Abstract
The collection the charged particles produced by the incident radiation on a detector requires a time interval. If this time interval is not sufficiently short compared with the peaking time of the amplifier, a loss in the recovered signal amplitude occurs. Another major constraint on the throughput of modern x or gamma-ray spectrometers is the time required for the subsequent the pulse processing by the electronics. Two above-mentioned limitations are cause of counting losses resulting from the dead time and the pile-up. The pulse pile-up is a common problem in x and gamma ray radiation detection systems. The pulses pile-up in spectroscopic analysis can cause significant errors. Therefore, inhibition of these pulses is a vital step. A way to reduce errors due to the pulse pile-up is a pile-up inspection circuitry (PUR). Such a circuit rejects some of the pulse pile-up. Therefore, this circuit leads to counting losses. Determination of these counting losses is an important problem. In this work, a new method is suggested for the determination of the pulse pile-up reject.Keywords
This publication has 6 references indexed in Scilit:
- Decaying source model: Alternative approach to determination of true counting rates at X and gamma ray counting systemsRadiation Measurements, 2013
- Determination of proper peaking time for Ultra-LEGe detectorRadiation Measurements, 2011
- A new method to compensate for counting losses due to system dead timeNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2009
- Time distortion of a Poisson process and its effect on experimental uncertaintyApplied Radiation and Isotopes, 1998
- Pulse Pile-Up, Dead Time, Derandomization, and Count Rate Capability in Scintillation Gamma CamerasIEEE Transactions on Nuclear Science, 1985
- Pile‐up rejection: Limitations and corrections for residual errors in energy‐dispersive spectrometersX-Ray Spectrometry, 1977