Technique for generating submicrometer ultra high molecular weight polyethylene particles

Authors:

Abstract
Ultra high molecular weight polyethylene wear debris is believed to have a major role in aseptic loosening of prosthetic joints. In order to study the cellular and host response to this and other such particulate debris, a source of fine ultra high molecular weight polyethylene debris is needed. We have described a technique to fracture the GUR 4150 primary ultra high molecular weight polyethylene grain, which reproducibly generated particles less than 1 μm in size. Furthermore, the particle morphology was similar to that of ultra high molecular weight polyethylene particles generated in vivo and retrieved from interfacial tissues. The fabricatd polyethylene particls ranged from 0.1 to 33 μm in diameter, with a mean of 2.3 ± 0.2 μm. Sixty percent of the particles were smaller than 1 μm and 90% were smaller than 7 μm. Using filtration and sedimentation, it is possible to acquire finer particle fractions. These particles are currently being used for biological response studies.