Biodegradable Sleeves for Metal Implants to Prevent Implant‐Associated Infection: An Experimental In Vivo Study in Sheep
- 12 January 2012
- journal article
- research article
- Published by Wiley in Veterinary Surgery
- Vol. 41 (3) , 410-421
Abstract
Objective To evaluate biocompatibility of biodegradable sleeves containing antimicrobial agents, designed for local drug delivery to prevent implant-related infection. Study Design Synthetic polyester sleeves (a copolymer of glycolide, caprolactone, trimethylene carbonate, lactide) were cast as thin films. The antimicrobial agents incorporated in the sleeves included gentamicin sulfate, triclosan, or a combination of these drugs. Animals Adult sheep (n = 15). Methods Two limited contact dynamic compression plates (LC-DCP) with or without sleeves were implanted on tibiae (bilateral) of 15 sheep. Sleeves were placed over the plates before implantation. Beneath half of the plates, 5-mm drill hole defects were made in the near cortex. Samples were harvested 4 weeks later for histology and microradiography. Results Macroscopically, no irritation of bone or adjacent tissue was seen. Small remnants of sleeves were visible on histology, and positively correlated with the presence of macrophages and foreign body cells. Thick sections showed no difference between the test samples and controls in terms of fibrous capsule formation, periosteal remodeling, and defect remodeling. Inflammatory cells, macrophages, and foreign body cells were more prominent in sections with sleeves, but were not statistically significantly different from controls. Cell numbers were within normal physiologic limits normally seen as cellular response to foreign bodies consisting of polymers. Conclusion The normal healing response indicated that the biodegradable sleeves demonstrate tissue biocompatibility.Keywords
This publication has 57 references indexed in Scilit:
- T cell subset distributions following primary and secondary implantation at subcutaneous biomaterial implant sitesJournal of Biomedical Materials Research Part A, 2007
- Periosteal Remodeling at the Femoral Neck in Nonhuman PrimatesJournal of Bone and Mineral Research, 2006
- Intraoperative Handling and Wound Healing: Controlled Clinical Trial Comparing Coated VICRYL® Plus Antibacterial Suture (Coated Polyglactin 910 Suture with Triclosan) with Coated VICRYL® Suture (Coated Polyglactin 910 Suture)Surgical Infections, 2005
- Lysostaphin-Coated Catheters Eradicate Staphylococccus aureus Challenge and Block Surface ColonizationAntimicrobial Agents and Chemotherapy, 2004
- Adhesion of bacteria to stainless steel and silver-coated orthopedic external fixation pinsJournal of Biomedical Materials Research, 1997
- Inhibition of Bacterial Adhesion and Biofilm Formation by a Heparinized Hydrophilic PolymerASAIO Journal, 1995
- The Potential of Sheep for the Study of Osteopenia: Current Status and Comparison with Other Animal ModelsBone, 1995
- An examination of the occurrence of surgical wound infection following equine orthopaedic surgery (1981–1990)Equine Veterinary Journal, 1994
- Clinical significance of translocation.Gut, 1994
- Two processes of bone remodeling in plated intact femora: An experimental study in dogsJournal of Orthopaedic Research, 1993