Wavelength, temperature, and voltage dependent calibration of a nematic liquid crystal multispectral polarization generating device.
- 31 July 2007
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 46 (22) , 5539-5544
Abstract
Rapid calibration of liquid crystal variable retarder (LCVR) devices is critical for successful clinical implementation of a LC-based Mueller matrix imaging system being developed for noninvasisve skin cancer detection. For multispectral implementation of such a system, the effect of wavelength (λ), temperature (T), and voltage (V) on the retardance (δ) required to generate each desired polarization state needs to be clearly understood. Calibration involves quantifying this interdependence such that for a given set of system input variables , the appropriate voltage is applied across a LC cell to generate a particular retardance. This paper presents findings that elucidate the dependence of voltage, for a set retardance, on the aforementioned variables for a nematic LC cell: λ-dependence and T-dependence. Additionally, an empirically derived model is presented that enables initial voltage calibration of retardance for any desired input wavelength within the calibration range of .
Keywords
This publication has 4 references indexed in Scilit:
- Optimized Mueller polarimeter with liquid crystalsOptics Letters, 2003
- Development and calibration of an automated Mueller matrix polarization imaging systemJournal of Biomedical Optics, 2002
- General and self-consistent method for the calibration of polarization modulators, polarimeters, and Mueller-matrix ellipsometersApplied Optics, 1999
- Birefringence measurements of liquid crystalsApplied Optics, 1984