The impact of MR‐based attenuation correction in spinal cord FDG‐PET/MR imaging for neurological studies
Open Access
- 9 August 2021
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 48 (10) , 5924-5934
Abstract
Purpose: Positron emission tomography (PET) attenuation correction (AC) in positron emission tomography‐magnetic resonance (PET/MR) scanners constitutes a critical and barely explored issue in spinal cord investigation, mainly due to the limitations in accounting for highly attenuating bone structures which surround the spinal canal. Our study aims at evaluating the clinical suitability of MR‐driven AC (MRAC) for 18‐fluorodeoxy‐glucose positron emission tomography (18F‐FDG‐PET) in spinal cord.Methods: Thirty‐six patients, undergoing positron emission tomography‐computed tomography (PET/CT) and PET/MR in the same session for oncological examination, were retrospectively analyzed.For each patient, raw PET data from PET/MR scanner were reconstructed with 4‐ and 5‐class MRAC maps, generated by hybrid PET/MR system (PET_MRAC4 and PET_MRAC5, respectively, where PET_MRAC is PET images reconstructed using MR‐based attenuation correction map), and an AC map derived from CT data after a custom co‐registration pipeline (PET_rCTAC, where PET_rCTAC is PET images reconstructed using CT‐based attenuation correction map), which served as reference. Mean PET standardized uptake values () were extracted from the three reconstructed PET images by regions of interest (ROIs) identified on T2‐weighted MRI, in the spinal cord, lumbar cerebrospinal fluid (CSF), and vertebral marrow at five levels (C2, C5, T6, T12, and L3). values from PET_MRAC4 and PET_MRAC5 were compared with each other and with the reference by means of paired t‐test, and correlated using Pearson's correlation (r) to assess their consistency. Cohen's d was calculated to assess the magnitude of differences between PET images.Results: values from PET_MRAC4 were lower than those from PET_MRAC5 in almost all analyzed ROIs, with a mean difference ranging from 0.03 to 0.26 (statistically significant in the vertebral marrow at C2 and C5, spinal cord at T6 and T2, and CSF at L3). This was also confirmed by the effect size, with highest values at low spinal levels (d = 0.45 at T12 in spinal cord, d = 0.95 at L3 in CSF). values from PET_MRAC4 and PET_MRAC5 showed a very good correlation (0.81 < r < 0.97, p < 0.05) in all spinal ROIs. Underestimation of between PET_MRAC4 and PET_rCTAC was observed at each level, with a mean difference ranging from 0.02 to 0.32 (statistically significant in the vertebral marrow at C2 and T6, and CSF at L3). Although PET_MRAC5 underestimates PET_rCTAC (mean difference ranging from 0.02 to 0.3), an overall decrease in effect size could be observed for PET_MRAC5, mainly at lower spinal levels (T12, L3). from both PET_MRAC4 and PET_MRAC5 methods showed r value from good to very good with respect to PET_rCTAC (0.67 < r < 0.9 and 0.73 < r < 0.94, p < 0.05, respectively).Conclusions: Our results showed that neglecting bones in AC can underestimate the FDG uptake measurement of the spinal cord. The inclusion of bones in MRAC is far from negligible and improves the AC in spinal cord, mainly at low spinal levels. Therefore, care must be taken in the spinal canal region, and the use of AC map reconstruction methods accounting for bone structures could be beneficial.Keywords
This publication has 60 references indexed in Scilit:
- Structural and Metabolic Features of Two Different Variants of Multiple Sclerosis: A PET/MRI StudyJournal of Neuroimaging, 2012
- Comparison of Segmentation-Based Attenuation Correction Methods for PET/MRI: Evaluation of Bone and Liver Standardized Uptake Value with Oncologic PET/CT DataJournal of Nuclear Medicine, 2012
- First Clinical Experience with Integrated Whole-Body PET/MR: Comparison to PET/CT in Patients with Oncologic DiagnosesJournal of Nuclear Medicine, 2012
- Effects of rigid and non-rigid image registration on test-retest variability of quantitative [18F]FDG PET/CT studiesEJNMMI Research, 2012
- Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR ScannerJournal of Nuclear Medicine, 2011
- A Systematic Review of the Factors Affecting Accuracy of SUV MeasurementsAmerican Journal of Roentgenology, 2010
- Tissue Classification as a Potential Approach for Attenuation Correction in Whole-Body PET/MRI: Evaluation with PET/CT DataJournal of Nuclear Medicine, 2009
- Towards quantitative PET/MRI: a review of MR-based attenuation correction techniquesEuropean Journal of Nuclear Medicine and Molecular Imaging, 2008
- Life of the mind: The interface of psychopharmaceuticals, domestic economies, and social abandonmentAmerican Ethnologist, 2004
- Comparing correlated correlation coefficients.Psychological Bulletin, 1992