Abstract
Three-dimensional vector imaging of bias-induced displacements of surfaces of ionically conductive materials using electrochemical strain microscopy (ESM) is demonstrated for model polycrystalline LiCoO2 surface. We demonstrate that resonance enhanced imaging using band excitation detection can be performed both for out-of-plane and in-plane response components at flexural and torsional resonances of the cantilever, respectively. The image formation mechanism in vector ESM is analyzed and relationship between measured signal and grain orientation is discussed.

This publication has 31 references indexed in Scilit: