Effect of mid-scan breathing changes on quality of 4DCT using a commercial phase-based sorting algorithm
Authors:
- Parag J. ParikhPPParag J. Parikh
- 28 April 2011
- journal article
- Published by Wiley in Medical Physics
- Vol. 38 (5) , 2430-2438
Abstract
Though it is known that irregular breathing can introduce artifacts in commercial 4DCT, this has not been systematically explored. The purpose of this study is to investigate the effect of variations in basic parameters of the breathing wave on 4DCT imaging quality. A four-dimensional motion platform holding an acrylic sphere was scanned while moving in a trajectory modeled from a lung cancer patient. A bellows device was used as a respiratory surrogate, and the images were sorted by a commercial phase-based sorting algorithm. Motion during the first half of the scan was produced at a baseline trajectory with a consistent frequency and amplitude of 15 breaths per minute and 1 cm, peak to peak. The two parameters were then varied mid-scan to new frequency and amplitude values, with frequencies ranging from 7.5 to 22 bpm and amplitudes ranging from 0.5 to 1.5 cm. Image sets representing four respiratory phases were contoured. Each set was analyzed to compare centroid displacement, density homogeneity, and volumetric and geometric distortions of the imaged sphere. Undercoverage of the target ITV and overcoverage of healthy tissue was also evaluated. Changes in amplitude of 25% or more, with or without changes in frequency, consistently caused measurable distortions in shape, position, and density of the imaged sphere. Frequency changes over 50% showed a similar trend. This study suggests that basic breathing statistics can be used to quickly assess the quality of a 4DCT scan prior to image reconstruction. Such information can help give indication of the proper course of action when irregular breathing patterns are observed during CT scanning.Keywords
This publication has 47 references indexed in Scilit:
- Retrospective Analysis of Artifacts in Four-Dimensional CT Images of 50 Abdominal and Thoracic Radiotherapy PatientsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Variability of Four-Dimensional Computed Tomography Patient ModelsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Can audio coached 4D CT emulate free breathing during the treatment course?Acta Oncologica, 2008
- Effect of Novel Amplitude/Phase Binning Algorithm on Commercial Four-Dimensional Computed Tomography QualityInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Phase and amplitude binning for 4D-CT imagingPhysics in Medicine & Biology, 2007
- Four-Dimensional Imaging and Treatment Planning of Moving Targets2007
- Investigation of patient, tumour and treatment variables affecting residual motion for respiratory-gated radiotherapyPhysics in Medicine & Biology, 2006
- The impact of tumor motion upon CT image integrity and target delineationMedical Physics, 2004
- Treatment plan optimization incorporating respiratory motionMedical Physics, 2004
- Respiration‐correlated spiral CT: A method of measuring respiratory‐induced anatomic motion for radiation treatment planningMedical Physics, 2002