Entrance surface dose measurements using a small OSL dosimeter with a computed tomography scanner having 320 rows of detectors
Authors:
- 24 June 2016
- journal article
- research article
- Published by Springer Nature in Radiological Physics and Technology
- Vol. 10 (1) , 49-59
Abstract
Entrance surface dose (ESD) measurements are important in X-ray computed tomography (CT) for examination, but in clinical settings it is difficult to measure ESDs because of a lack of suitable dosimeters. We focus on the capability of a small optically stimulated luminescence (OSL) dosimeter. The aim of this study is to propose a practical method for using an OSL dosimeter to measure the ESD when performing a CT examination. The small OSL dosimeter has an outer width of 10 mm; it is assumed that a partial dose may be measured because the slice thickness and helical pitch can be set to various values. To verify our method, we used a CT scanner having 320 rows of detectors and checked the consistencies of the ESDs measured using OSL dosimeters by comparing them with those measured using Gafchromic™ films. The films were calibrated using an ionization chamber on the basis of half-value layer estimation. On the other hand, the OSL dosimeter was appropriately calibrated using a practical calibration curve previously proposed by our group. The ESDs measured using the OSL dosimeters were in good agreement with the reference ESDs from the Gafchromic™ films. Using these data, we also estimated the uncertainty of ESDs measured with small OSL dosimeters. We concluded that a small OSL dosimeter can be considered suitable for measuring the ESD with an uncertainty of 30 % during CT examinations in which pitch factors below 1.000 are applied.Keywords
This publication has 28 references indexed in Scilit:
- Low-dose protocol of the spiral CT in orthodontics: comparative evaluation of entrance skin dose with traditional X-ray techniquesProgress in Orthodontics, 2013
- Radiation dose reduction by using low dose CT protocol of thoraxRadiation Measurements, 2013
- Cancer risk in 680 000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million AustraliansBMJ, 2013
- Dose Reduction to Anterior Surfaces With Organ-Based Tube-Current Modulation: Evaluation of Performance in a Phantom StudyAmerican Journal of Roentgenology, 2011
- CT Dose Index and Patient Dose: They AreNotthe Same ThingRadiology, 2011
- Estimating Effective Dose for CT Using Dose–Length Product Compared With Using Organ Doses: Consequences of Adopting International Commission on Radiological Protection Publication 103 or Dual-Energy ScanningAmerican Journal of Roentgenology, 2010
- Reference radiochromic film dosimetry in kilovoltage photon beams during CBCT image acquisitionMedical Physics, 2010
- Digital radiography: The balance between image quality and required radiation doseEuropean Journal of Radiology, 2009
- Principles of CT: Radiation Dose and Image QualityJournal of Nuclear Medicine Technology, 2007
- Photon mass attenuation and energy-absorption coefficientsThe International Journal of Applied Radiation and Isotopes, 1982