Xurography actuated valving for centrifugal flow control
- 9 August 2016
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Lab on a Chip
- Vol. 16 (18) , 3454-3459
Abstract
We introduce a novel instrument controlled valving scheme for centrifugal platforms which is based upon xurography. In a first approach, which is akin to previously presented event-triggered flow control, the valves are composed of a pneumatic chamber sealed by a dissolvable film (DF) and by a pierceable membrane. Liquid is initially prevented from wetting the DF by the counter pressure of a trapped gas. Via a channel, this pocket is pneumatically connected to a vent, sealed by the pierceable membrane, located on the top surface of the disc. By scouring the top surface of the disc, along a pre-defined track by a robotic knife-cutter, the trapped gas is released and so the liquid can wet and disintegrate the DF. In order to automate assay protocols without the need to integrate DFs, we extend this xurography-based flow control concept by selective venting of chambers subjected to pneumatic over-pressure or vacuum suction. Unlike most instrument controlled flow-control mechanisms, in this approach to valve actuation can occur during disc rotation. To demonstrate the potential of this flow control approach, we designed a disc architecture to automate the liquid handling as the backbone of a biplex liver assay panel. We demonstrate valve actuation during rotation, using the robotic arm, using this disc with visualisation via dyed water. We then demonstrate the biplex liver assay, using calibration reagent, by stopping the disc and manually piercing the membrane to actuate the same valves.Keywords
Funding Information
- Seventh Framework Programme (FP7-KBBE-2013-7-613908-DECATHLON)
- Science Foundation Ireland (10/CE/B1821, 12/IA/1567, 12/RC/2289-INSIGHT)
This publication has 17 references indexed in Scilit:
- Centrifugal microfluidic platforms: advanced unit operations and applicationsChemical Society Reviews, 2015
- Paper imbibition for timing of multi-step liquid handling protocols on event-triggered centrifugal microfluidic lab-on-a-disc platformsRSC Advances, 2014
- Event-triggered logical flow control for comprehensive process integration of multi-step assays on centrifugal microfluidic platformsLab on a Chip, 2014
- Vacuum/Compression Valving (VCV) Using Parrafin-Wax on a Centrifugal Microfluidic CD PlatformPLOS ONE, 2013
- Centrifugo-pneumatic valving utilizing dissolvable filmsLab on a Chip, 2012
- Optically addressable single-use microfluidic valves by laser printer lithographyLab on a Chip, 2010
- Centrifugal microfluidics for biomedical applicationsLab on a Chip, 2010
- The centrifugal microfluidic Bio-Disk platformJournal of Micromechanics and Microengineering, 2007
- One-step pathogen specific DNA extraction from whole blood on a centrifugal microfluidic deviceLab on a Chip, 2007
- LAB ON A CDAnnual Review of Biomedical Engineering, 2006