TFTR multichannel neutron collimator
Authors:
- 1 October 1990
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 61 (10) , 3163-3165
Abstract
The TFTR multichannel neutron collimator provides time‐resolved neutron emission profiles from ten vertical chords spaced in major radius across the vacuum vessel. An in situ calibration of the system has been made using a 120 mCi 252Cf source scanned inside the vacuum vessel. A mapping of the effective neutron acceptance aperture was performed and showed reasonable agreement with the optical predictions testifying to the effectiveness of the interstitial shielding material. Crosstalk between channels was evaluated and found to be negligible. Periodic renormalization of the system has shown a standard deviation among the sensitivities of only 2.0% over the first seven months of operation.Keywords
This publication has 3 references indexed in Scilit:
- TFTR epithermal neutron detector system: Recalibration and effect of nonisotropic neutron emissionReview of Scientific Instruments, 1988
- Collimated ZnS (Ag)-detector array for Tokamak Fusion Test Reactor neutron source strength radial profile measurementsReview of Scientific Instruments, 1985
- ZnS(Ag) Phosphor Mixtures for Neutron Scintillation CountingReview of Scientific Instruments, 1955