Temperature monitoring in fat with MRI
Open Access
- 7 June 2000
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 43 (6) , 901-904
Abstract
The aim of the study was to test the hypothesis that fast spin echo T1‐weighted images can be used to quantify the temperature in fat during thermal therapy in vivo. An MR compatible positioning device was used to manipulate focused ultrasound transducers in an MRI scanner. This system was used to sonicate fat tissue around the kidneys of 12 rabbits at various power levels for 10 to 20 sec. The scan parameters of T1‐weighted fast spin echo (FSE) sequence were varied to optimize signal intensity characteristics while maintaining short scan times. An invasive optical probe was used to calibrate the temperature related signal intensity changes. For the T1‐weighted FSE sequence, the signal intensity decreased with the temperature elevation at the rate of 0.97 ± 0.02%/oC. The single focused transducer produced a contrast‐to‐noise ratio more than 10 at power levels below the tissue damage threshold. The signal intensity was linearly dependent on the power, despite the measured temperatures being well above the coagulation threshold. This study demonstrates that T1‐weighted FSE MRI sequences can be used to quantify the temperature elevation in fat in vivo during short focused ultrasound exposures. This can be very important for breast tumor surgery, fat ablation, and for treating deep seated tumors through superficial fat layers. Magn Reson Med 43:901–904, 2000.Keywords
This publication has 16 references indexed in Scilit:
- Optimization of chemical shift selective suppression of fatMagnetic Resonance in Medicine, 1998
- Design and evaluation of a feedback based phased array system for ultrasound surgeryIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1998
- Time and temperature dependence of MR parameters during thermal coagulation of ex vivo rabbit muscleMagnetic Resonance in Medicine, 1998
- Thermal effects of focused ultrasound on the brain: determination with MR imaging.Radiology, 1997
- A clinical, noninvasive, MR imaging-monitored ultrasound surgery methodRadioGraphics, 1996
- Focused US system for MR imaging-guided tumor ablation.Radiology, 1995
- The Usefulness of a Contrast Agent and Gradient-Recalled Acquisition in a Steady-State Imaging Sequence for Magnetic Resonance Imaging-Guided Noninvasive Ultrasound SurgeryInvestigative Radiology, 1994
- Noninvasive temperature imaging using diffusion MRIMagnetic Resonance in Medicine, 1991
- Concentric-Ring and Sector-Vortex Phased-Array Applicators for Ultrasound HyperthermiaIEEE Transactions on Microwave Theory and Techniques, 1986
- Applications of NMR Imaging in Hyperthermia: An Evaluation of the Potential for Localized Tissue Heating and Noninvasive Temperature MonitoringIEEE Transactions on Biomedical Engineering, 1984