Technical Note: Characterization and correction of gradient nonlinearity induced distortion on a 1.0 T open bore MR‐SIM
- 22 September 2015
- journal article
- Published by Wiley in Medical Physics
- Vol. 42 (10) , 5955-5960
Abstract
Purpose: Distortions in magnetic resonance imaging (MRI) compromise spatial fidelity, potentially impacting delineation and dose calculation. We characterized 2D and 3D large field of view (FOV), sequence‐independent distortion at various positions in a 1.0 T high‐field open MR simulator (MR‐SIM) to implement correction maps for MRI treatment planning.Methods: A 36 × 43 × 2 cm3 phantom with 255 known landmarks (∼1 mm3) was scanned using 1.0 T high‐field open MR‐SIM at isocenter in the transverse, sagittal, and coronal axes, and a 465 × 350 × 168 mm3 3D phantom was scanned by stepping in the superior‐inferior direction in three overlapping positions to achieve a total 465 × 350 × 400 mm3 sampled FOV yielding >13 800 landmarks (3D Gradient‐Echo, TE/TR/α = 5.54 ms/30 ms/28°, voxel size = 1 × 1 × 2 mm3). A binary template (reference) was generated from a phantom schematic. An automated program converted MR images to binary via masking, thresholding, and testing for connectivity to identify landmarks. Distortion maps were generated by centroid mapping. Images were corrected via warping with inverse distortion maps, and temporal stability was assessed.Results: Over the sampled FOV, non‐negligible residual gradient distortions existed as close as 9.5 cm from isocenter, with a maximum distortion of 7.4 mm as close as 23 cm from isocenter. Over six months, average gradient distortions were −0.07 ± 1.10 mm and 0.10 ± 1.10 mm in the x and y directions for the transverse plane, 0.03 ± 0.64 and −0.09 ± 0.70 mm in the sagittal plane, and 0.4 ± 1.16 and 0.04 ± 0.40 mm in the coronal plane. After implementing 3D correction maps, distortions were reduced to <1 pixel width (1 mm) for all voxels up to 25 cm from magnet isocenter.Conclusions: Inherent distortion due to gradient nonlinearity was found to be non‐negligible even with vendor corrections applied, and further corrections are required to obtain 1 mm accuracy for large FOVs. Statistical analysis of temporal stability shows that sequence independent distortion maps are consistent within six months of characterization.Keywords
Funding Information
- National Institutes of Health (R01EB016079)
- Philips
- Henry Ford Health System
- National Institutes of Health (NIH) (R01EB016079)
- Philips
- Henry Ford Health System
- National Institutes of Health (R01EB016079)
- Philips
- Henry Ford Health System
This publication has 30 references indexed in Scilit:
- Investigation of a method for generating synthetic CT models from MRI scans of the head and neck for radiation therapyPhysics in Medicine & Biology, 2013
- Systematisation of spatial uncertainties for comparison between a MR and a CT-based radiotherapy workflow for prostate treatmentsRadiation Oncology, 2009
- A study on the magnetic resonance imaging (MRI)-based radiation treatment planning of intracranial lesionsPhysics in Medicine & Biology, 2008
- A complete distortion correction for MR images: I. Gradient warp correctionPhysics in Medicine & Biology, 2005
- Geometric distortion in clinical MRI systemsMagnetic Resonance Imaging, 2004
- A novel phantom and method for comprehensive 3-dimensional measurement and correction of geometric distortion in magnetic resonance imagingMagnetic Resonance Imaging, 2004
- CT and MRI derived source localization error in a custom prostate phantom using automated image coregistrationMedical Physics, 2001
- Correction for geometric distortion in echo planar images from B0 field variationsMagnetic Resonance in Medicine, 1995
- A technique for accurate magnetic resonance imaging in the presence of field inhomogeneitiesIEEE Transactions on Medical Imaging, 1992
- Analysis of machine-dependent and object-induced geometric distortion in 2DFT MR imagingMagnetic Resonance Imaging, 1992