A new second-generation supraglottic airway device (Ambu® AuraGain•) versus intubating laryngeal mask airway as conduits for blind intubation – A prospective, randomised trial
Authors:
Open Access
- 1 January 2019
- journal article
- research article
- Published by Medknow in Indian Journal of Anaesthesia
- Vol. 63 (7) , 558-564
Abstract
Background and Aims: The Ambu® AuraGain• is a new single-use supraglottic airway device with gastric channel, designed to facilitate intubation. The study aimed to assess the success rates of proper placement and intubation using Ambu® AuraGain• compared with intubating laryngeal mask airway?(ILMA). Methods: One hundred and twenty?patients?(18–60?years) were enrolled into this prospective, randomised, comparative study. After inducing general anaesthesia, appropriate size ILMA?(group?I)/Ambu® AuraGain•?(group?A) was placed as per the manufacturer's recommendations and correct placement was confirmed. Appropriate size endotracheal tube was passed through the device. The success rate of insertion and intubation, number of attempts, Cormack–Lehane grading before insertion and haemodynamics were recorded. Data were analysed using Mantel–Haenszel Chi-square test, Student's t-test and Fisher's exact test. Results: Demographic and airway parameters were uniformly distributed in both the groups. The success rate for insertion was 100% in both devices. The success rate for intubation was 96.6%?(58/60) in group?I and 36.6% in group?A?(P?7.62?±?0.75?cm had higher successful intubation compared with patients with mean thyromental distance?P?=?0.014). Cormack–Lehane grading did not correlate with intubation attempts or success rate in group?I?(P?=?0.45), whereas in group?A the rate of successful blind intubation with Cormack–Lehane grade?1 was 50%?(19/38). Conclusion: Both devices have 100% insertion success, though Ambu® AuraGain• has lower success rate for facilitating intubation compared with ILMA.Keywords
This publication has 14 references indexed in Scilit:
- Difficult Airway Society 2015 guidelines for management of unanticipated difficult intubation in adultsBritish Journal of Anaesthesia, 2015
- Comparison of supraglottic devices i-gel ® and LMA Fastrach ® as conduit for endotracheal intubationIndian Journal of Anaesthesia, 2014
- Effects of tracheal tube orientation on the success of intubation through an intubating laryngeal mask airway: study in Mallampati class 3 or 4 patientsBritish Journal of Anaesthesia, 2009
- Use of the Intubating LMA-Fastrach™ in 254 Patients with Difficult-to-manage AirwaysAnesthesiology, 2001
- A new practical classification of laryngeal viewAnaesthesia, 2000
- The Intubating Laryngeal Mask Airway With and Without Fiberoptic GuidanceAnesthesia & Analgesia, 1999
- The intubating laryngeal mask. II: A preliminary clinical report of a new means of intubating the tracheaBritish Journal of Anaesthesia, 1997
- What Is the Best Way to Determine Oropharyngeal Classification and Mandihular Space Length to Predict Difficult Laryngoscopy?Anesthesiology, 1994
- Anaesthetic implications of temporomandibular joint diseaseCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1992
- A clinical sign to predict difficult tracheal intubation; a prospective studyCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1985