The primary objective of this study was to establish the safety, efficacy, infection rate, and patency of the Vectra graft (polyetherurethaneurea) for dialysis access in patients diagnosed with human immunodeficiency virus (HIV) and end-stage renal disease. The Vectra graft has a unique self-sealing property; therefore we hypothesize that these patients will have fewer infections. A Vectra graft was implanted in 30 consecutive HIV-positive patients without sufficient veins for an autologous fistula. These surgeries were carried out over a 2.5-year period. Primary graft patency was 42% at 12 months and 3 (10%) of the grafts developed infection. This rate of graft infection was less (10% vs 45%) than both our prior experience and published reports using polytetrafluorothene bridge grafts. The unique self-sealing property of the Vectra graft minimizes the development of perigraft hematoma with repetitive needle cannulation and in the immunosuppressed HIV-positive patient, may account for the observed decrease in dialysis access infection.Keywords
This publication has 9 references indexed in Scilit:
- Human Immunodeficiency Virus Infection in End‐Stage Renal Disease PatientsSeminars in Dialysis, 2003
- Special issues in the management of patients with ESRD and HIV infectionAmerican Journal of Kidney Diseases, 2003
- Multicenter evaluation of a polyurethaneurea vascular access graft as compared with the expanded polytetrafluoroethylene vascular access graft in hemodialysis applicationsJournal of Vascular Surgery, 2001
- Infectious complications of the hemodialysis accessKidney International, 2001
- Tunneled-Cuffed Catheter Associated Infections in Hemodialysis Patients Who Are Seropositive for the Human Immunodeficiency VirusJournal of the American Society of Nephrology, 2000
- Polytetrafluoroethylene Hemoaccess Site InfectionsASAIO Journal, 2000
- Hemodialysis access: Influence of the human immunodeficiency virus on patency and infection ratesJournal of Vascular Surgery, 1999
- The influence of human immunodeficiency virus infection and intravenous drug abuse on complications of hemodialysis access surgeryJournal of Vascular Surgery, 1992
- Mortality in Dialysis Patients: Analysis of the Causes of DeathAmerican Journal of Kidney Diseases, 1991