Quantification and Detectability of In-Stent Stenosis with CT Angiography and MR Angiography in Arterial Stents In Vitro

Authors:

Abstract
OBJECTIVE. The purpose of this study was to compare CT angiography (CTA) and MR angiography (MRA) for the detectability of 75% and 95% stenoses in phantoms using six different stents.MATERIALS AND METHODS. Six different stents (Expander, Hemobahn, SelfX, Smart, Symphony, and Wallstent) were inserted into tubes filled with contrast agent (ioversol or gadoteric acid). To mimic stenoses of 75% and 95% of the patent lumen, 8-mm-diameter nylon cylinders were bored in the central axis (2 mm and 4 mm, respectively) and placed into the stent lumen. Intensity profiles across stenoses on 2-mm coronal reformatted sections of CTA or MRA were compared, and the detectability of the residual lumen was assessed using a subjective score.RESULTS. CTA showed relative in-stent signal attenuation for the in-stent stenoses of the tested stents ranging from 75% to 100% of the signal intensity of the control. SelfX and Symphony showed further shading of the residual lumen due to beam-hardening artifacts. Overestimation of stenos...

This publication has 25 references indexed in Scilit: