Photon-noise-limited operation of intensified CCD cameras
Authors:
- 1 August 1997
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 36 (22) , 5288-5297
Abstract
Intensified CCD cameras are increasingly being used in quantitative applications, which requires not only a greater understanding of their operation but also more detailed modeling to predict their performance more accurately. We have developed a model based on photon-noise-limited operation that incorporates the effects of the point spread function of the intensifier on signal-to-noise ratio. These effects are absent in other models, which renders them inadequate to model the camera performance properly. Calculations of noise-equivalent irradiance with our model are shown to be in good agreement with experimental results presented for two Xybion intensified cameras, Models GEN-III IMC and NIR DCIC intensified cameras.Keywords
This publication has 4 references indexed in Scilit:
- A directional fast neutron detector using scintillating fibers and an intensified CCD camera systemNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1994
- An economical x-ray area detector based on an intensified CCD unitReview of Scientific Instruments, 1993
- Application of the intensified CCD to airglow and auroral measurementsApplied Optics, 1992
- Statistical performance of the intensified charged coupled deviceApplied Optics, 1986