Validated Method for the Quantification of Atractylenolide III in Different Processed Products of Rhizoma Atractylodes Macrocephalae

Authors:

Abstract
Introduction – Rhizoma Atractylodes Macrocephalae (RAM) contains several sesquiterpene compounds including atractylenolide III (AO‐III). This bioactive compound may be used as a chemical marker for the quality control of different processed RAM products.Objective – To develop and validate an RP‐HPLC method for the quantitative determination of AO‐III in RAM and in a variety of processed RAM products.Methodology – HPLC was carried out using a Kromssil C18 RP‐column eluted with methanol–water (70:30) at a flow rate of 1.0 mL/min and with UV detection at 220 nm. Full validation was performed using standard methods.Results – The linear range of AO‐III was 5–50 µg/mL; the regression equation was y = 10210x + 11194 (r = 0.9994) and the average recovery was 101.08% (RSD = 0.98%). The detection and quantification limits for AO‐III were determined to be 0.005 and 0.018 µg/mL at signal‐to‐noise ratios of approximately 3:1 and 10:1, respectively.Conclusion – The described HPLC method is appropriate for quality assurance and differentiation of AO‐III in RAM and different processed products.