An ion trap storage/time-of-flight mass spectrometer

Authors:

Abstract
An ion trap/time‐of‐flight (TOF) mass spectrometer combination has been developed in order to combine the storage capabilities of an ion trap with the speed and resolution of a time‐of‐flight device. The ion trap is an rf quadrupole trap which operates in the total storage mode, i.e., with the dc voltage=0 on the end caps and rf voltage on the ring electrode. The trap has an ion storage time of ≥2 s at an rf potential of 310 Vpp. The stored ions are ejected into the time‐of‐flight device using a −150 V dc pulse on the exit end cap which causes the ion trajectories in the trap to become unstable. The ions are mass analyzed using either a linear or reflectron TOF. In the linear mode the resolution is 240 while in the reflectron mode a resolution of 1300 at m/z 93 is achieved. The storage capabilities of this device may have important applications towards enhancing sensitivity, the study of very slow metastable decay, and photodissociation mass spectrometry and spectroscopy.

This publication has 28 references indexed in Scilit: