Dosimetric characterization of an encapsulated waterproof silicon carbide detector with UHDR electron and proton beams for FLASH radiotherapy
Open Access
- 16 October 2025
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 70 (20) , 205019
Abstract
Silicon carbide (SiC) detectors have been widely demonstrated to be suitable alternative detectors for dosimetry in FLASH radiotherapy, showing radiation hardness and dose-rate independence at the FLASH radiotherapy instantaneous dose rates (IDRs). However, the practical use of such devices in the preclinical/clinical environment still requires the development of special handy housing enabling the quality assurance (QA) measurements under the reference dosimetric conditions. A 10 μm thick, 4.5 mm2 area SiC detector produced by the STLab company was recently embedded at the INFN-Catania Division inside a plastic waterproof 15 mm diameter cylindrical housing. This encapsulated version of SiC (eSiC) allows the measurement of the dose in reference conditions and of the dose profiles in liquid/solid water phantoms for assuring high accuracy dosimetry QA procedures. Dosimetric characterizations were performed with both electron and proton beams at conventional and ultra-high dose rates (UHDR). A first experiment was carried out at the Centro Pisano for Flash Radiotherapy using UHDR 9 MeV electron beams to confirm the linearity of the charge response as a function of the dose per pulse after the encapsulation procedure. A linearity from 1.8 Gy/pulse up to about 12 Gy/pulse, corresponding to an IDR of 3 MGy s−1, was found. The percentage depth dose (PDD) distribution in water of 9 MeV electron beams was also measured and compared with the PDD measured with a Freiburg Physikalisch-Technische Werkstätten Dr. Pychlau GmbH (PTW) flash diamond detector, used as reference dosimeter. The eSiC detector was also tested with proton beams accelerated by the IBA Proteus 235 cyclotron at the Trento Proton Therapy facility. A response independence on the total delivered dose (1–30 Gy) and average dose rate (50–530 Gy s−1) was found using the UHDR 228 MeV proton beam available along the experimental beamline. The depth dose distribution measured with the eSiC within a liquid water phantom was successfully compared with the one simultaneously measured by the IBA PPC05 reference chamber, using 180 MeV clinical proton beams. The excellent results demonstrated that this first realized eSiC prototype can be used to accurately perform reference and relative dosimetry with UHDR electron and proton beams, contributing to support the clinical translation of FLASH radiotherapy.Keywords
Funding Information
- The National Plan for NRRP Complementary Investments, Project: AdvaNced Technologies for Human-centrEd Medicine (Project n. PNC0000003)
This publication has 39 references indexed in Scilit:
- Proton FLASH Radiotherapy for the Treatment of Symptomatic Bone MetastasesJAMA Oncology, 2023
- Ultrahigh dose rate pencil beam scanning proton dosimetry using ion chambers and a calorimeter in support of first in‐human FLASH clinical trialMedical Physics, 2022
- Ultra‐high dose rate dosimetry: Challenges and opportunities for FLASH radiation therapyMedical Physics, 2022
- Development of an ultra‐thin parallel plate ionization chamber for dosimetry in FLASH radiotherapyMedical Physics, 2022
- Design, realization, and characterization of a novel diamond detector prototype for FLASH radiotherapy dosimetryMedical Physics, 2022
- Spread‐out Bragg peak proton FLASH irradiation using a clinical synchrocyclotron: Proof of concept and ion chamber characterizationMedical Physics, 2021
- Commissioning of an ultra‐high dose rate pulsed electron beam medical LINAC for FLASH RT preclinical animal experiments and future clinical human protocolsMedical Physics, 2021
- Ion collection efficiency of ionization chambers in ultra‐high dose‐per‐pulse electron beamsMedical Physics, 2020
- High dose‐per‐pulse electron beam dosimetry: Usability and dose‐rate independence of EBT3 Gafchromic filmsMedical Physics, 2017
- High dose‐per‐pulse electron beam dosimetry — A model to correct for the ion recombination in the Advanced Markus ionization chamberMedical Physics, 2017