A Rare Case of Adverse Reaction to Metal Debris in a Ceramic-on-Ceramic Total Hip Replacement
Open Access
- 8 September 2022
- journal article
- case report
- Published by MDPI AG in Journal of Functional Biomaterials
- Vol. 13 (3) , 145
Abstract
Adverse Reaction to Metal Debris (ARMD) is one of the most frequent complications after Total Hip Replacement (THR) and often a cause of surgical revision. This is true especially for implants with Metal-on-Metal (MoM) and Large Diameter Heads (LDHs), which are frequently used to improve stability and reduce the risk of dislocation. However, ARMD is not exclusive to MoM replacement, as it can also occur in other implants such Ceramic-on-Ceramic (CoC), Metal-on-Polyethylene (MoP), and Ceramic-on-Polyethylene (CoP). In these non-MoM implants, ARMD is not caused by the tribo-corrosion between bearing surfaces but, rather, by the fretting at the interface between neck and stem of dissimilar metals. A case of a severe ARMD that happened to a 73-year-old female patient with CoC bearing THR at the right hip is presented in this case report. In this case, the ARMD was misdiagnosed for over a year, resulting in the development of a massive pseudotumor. The treatment of choice was a two-stage revision with the implant of a hip megaprosthesis. After more than 2 years of follow-up, complete recovery of hip Range of Motion (ROM) and normalization of chromium and cobalt levels in blood and urine were achieved. Despite the relatively short follow-up period, this can be considered a successful treatment of a major and misdiagnosed ARMD in a non-MoM hip replacement.Keywords
This publication has 29 references indexed in Scilit:
- Corrosion and Fretting of a Modular Hip System: A Retrieval Analysis of 60 Rejuvenate StemsThe Journal of Arthroplasty, 2015
- Adverse tissue reaction to corrosion at the neck-stem junction after modular primary total hip arthroplastyOrthopaedics & Traumatology: Surgery & Research, 2015
- Adverse Reactions to Metal Debris in Metal-On-Polyethylene Total Hip Arthroplasty Using a Titanium-Molybdenum-Zirconium-Iron Alloy StemThe Journal of Arthroplasty, 2015
- Predictors of time to revision and clinical outcomes following revision of metal-on-metal hip replacements for adverse reaction to metal debrisThe Bone & Joint Journal, 2014
- Revision of Metal-on-Metal Hip Replacements and Resurfacings for Adverse Reaction to Metal Debris: A Systematic Review of OutcomesHIP International, 2014
- Fretting corrosion of CoCrMo and Ti6Al4V interfacesBiomaterials, 2012
- Large-Diameter Modular Metal-on-Metal Total Hip ArthroplastyThe Journal of Arthroplasty, 2012
- Pseudotumors in Association with Well-Functioning Metal-on-Metal Hip ProsthesesJournal of Bone and Joint Surgery, 2012
- Adverse reaction to metal debris following hip resurfacingThe Journal of Bone and Joint Surgery. British volume, 2011
- Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacementThe Journal of Bone and Joint Surgery. British volume, 2010