Towards a Customizable, SLA 3D-Printed Biliary Stent: Optimizing a Commercially Available Resin and Predicting Stent Behavior with Accurate In Silico Testing
Authors:
- Victoria CordistaVCVictoria Cordista
- Sagar PatelSPSagar Patel
- Rebecca LawsonRLRebecca Lawson
- Gunhee LeeGLGunhee Lee
- Morgan VerheyenMVMorgan Verheyen
- Ainsley WestbrookAWAinsley Westbrook
- Nathan SheltonNSNathan Shelton
- Prakriti SapkotaPrakriti Sapkota
- Isabella Zabala ValenciaIsabella Zabala Valencia
- Cynthia GaddamCynthia Gaddam
- Joanna ThomasJTJoanna Thomas
Open Access
Abstract
Inflammation of the bile ducts and surrounding tissues can impede bile flow from the liver into the intestines. If this occurs, a plastic or self-expanding metal (SEM) stent is placed to restore bile drainage. United States (US) Food and Drug Administration (FDA)-approved plastic biliary stents are less expensive than SEMs but have limited patency and can occlude bile flow if placed spanning a duct juncture. Recently, we investigated the effects of variations to post-processing and autoclaving on a commercially available stereolithography (SLA) resin in an effort to produce a suitable material for use in a biliary stent, an FDA Class II medical device. We tested six variations from the manufacturer’s recommended post-processing and found that tripling the isopropanol (IPA) wash time to 60 min and reducing the time and temperature of the UV cure to 10 min at 40 °C, followed by a 30 min gravity autoclave cycle, yielded a polymer that was flexible and non-cytotoxic. In turn, we designed and fabricated customizable, SLA 3D-printed polymeric biliary stents that permit bile flow at a duct juncture and can be deployed via catheter. Next, we generated an in silico stent 3-point bend test to predict displacements and peak stresses in the stent designs. We confirmed our simulation accuracy with experimental data from 3-point bend tests on SLA 3D-printed stents. Unfortunately, our 3-point bend test simulation indicates that, when bent to the degree needed for placement via catheter (~30°), the peak stress the stents are predicted to experience would exceed the yield stress of the polymer. Thus, the risk of permanent deformation or damage during placement via catheter to a stent printed and post-processed as we have described would be significant. Moving forward, we will test alternative resins and post-processing parameters that have increased elasticity but would still be compatible with use in a Class II medical device.Keywords
This publication has 64 references indexed in Scilit:
- Biosynthetic Plastics as Tunable Elastic and Visible Stent with Shape‐Memory to Treat Biliary StrictureAdvanced Science, 2023
- Challenges and Opportunities in 3D Printing of Biodegradable Medical Devices by Emerging Photopolymerization TechniquesAdvanced Functional Materials, 2022
- The Exploration of a Novel Biodegradable Drug-Eluting Biliary Stent: Preliminary WorkCardioVascular and Interventional Radiology, 2021
- Common bile duct pressure in patients with and without cholelithiasis: A case‐control studyJournal of Hepato-Biliary-Pancreatic Sciences, 2021
- Effects of different sterilization processes on the properties of a novel 3D‐printed polycaprolactone stentPolymers for Advanced Technologies, 2018
- Efficacy of a Metallic Stent Covered with a Paclitaxel-Incorporated Membrane Versus a Covered Metal Stent for Malignant Biliary Obstruction: A Prospective Comparative StudyDigestive Diseases and Sciences, 2012
- Structural mechanics of semicrystalline polymers prior to the yield point: a reviewJournal of Materials Science, 2012
- Investigation into double yield points in polyethylenePolymer, 1992
- Double yield point in polyethylene under tensile loadingJournal of Materials Science Letters, 1990
- Pressure measurements in the biliary and pancreatic duct systems in controls and in patients with gallstones, previous cholecystectomy, or common bile duct stonesGastroenterology, 1979