Validation and Feasibility of Intraoperative Three-Dimensional Transesophageal Echocardiographic Cardiac Output
Authors:
- William C. CulpWCWilliam C. Culp
- Timothy R. BallTBTimothy R. Ball
- Christopher J. BurnettCBChristopher J. Burnett
- 1 November 2007
- journal article
- research article
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 105 (5) , 1219-1223
Abstract
In this pilot study, we attempted to validate three-dimensional transesophageal echocardiography (3DTEE) cardiac output and assess its feasibility intraoperatively. Twenty patients undergoing cardiac surgery underwent simultaneous cardiac output determinations during the clinically stable prebypass period by 3DTEE and thermodilution. The correlation coefficient between cardiac output measured by the two methods was 0.86. The 3DTEE mean bias was 0.27 L/min, limits of agreement -1.64 to 2.17 L/min (approximately +/-35%). Three-dimensional data acquisition averaged 43 s; postprocessing took 7 min. Three-dimensional TEE can measure cardiac output and is feasible perioperatively. Measurements have good correlation with thermodilution, though with a significant bias and wide limits of agreement.Keywords
This publication has 17 references indexed in Scilit:
- Cardiac Output Measurement During Infrarenal Aortic Surgery: Echo-Esophageal Doppler Versus Thermodilution CatheterJournal of Cardiothoracic and Vascular Anesthesia, 2006
- Real-Time Transthoracic Three-Dimensional Echocardiographic Assessment of Left Ventricular Volume and Ejection Fraction in Congenital Heart DiseaseJournal of the American Society of Echocardiography, 2006
- Hemodynamic Measurements After Cardiac Surgery: Transesophageal Doppler Versus Pulmonary Artery CatheterJournal of Cardiothoracic and Vascular Anesthesia, 2005
- Recommendations for Chamber Quantification: A Report from the American Society of Echocardiography’s Guidelines and Standards Committee and the Chamber Quantification Writing Group, Developed in Conjunction with the European Association of Echocardiography, a Branch of the European Society of CardiologyJournal of the American Society of Echocardiography, 2005
- Clinical Features of Complications From Transesophageal Echocardiography: A Single-Center Case Series of 10,000 Consecutive ExaminationsJournal of the American Society of Echocardiography, 2005
- Comparison of left ventricular volumes and ejection fractions measured by three-dimensional echocardiography versus by two-dimensional echocardiography and cardiac magnetic resonance in patients with various cardiomyopathiesThe American Journal of Cardiology, 2005
- Transoesophageal echocardiography is unreliable for cardiac output assessment after cardiac surgery compared with thermodilution*Anaesthesia, 2004
- Rapid Quantification of Left Ventricular Function and Mass Using Transoesophageal Three-dimensional Echocardiography: Validation of a Method that uses Long-axis Cutplanes*European Heart Journal - Cardiovascular Imaging, 2000
- Intraoperative Determination of Cardiac Output Using Multiplane Transesophageal EchocardiographyAnesthesiology, 1998
- Comparison of cardiac output measurement techniques: thermodilution, Doppler, GO2‐rebreathing and the direct Fick methodActa Anaesthesiologica Scandinavica, 1995