Biologic Response to Polymer-coated Stents: In Vitro Analysis and Results in an Iliac Artery Sheep Model
Authors:
- Joerg LahannJLJoerg Lahann
- Pascal NiggemannPNPascal Niggemann
- Bernd KlosterhalfenBKBernd Klosterhalfen
- Julius MeyerJMJulius Meyer
- Arthur KulischAKArthur Kulisch
- Doris KleeDKDoris Klee
- Rolf W. GüntherRGRolf W. Günther
- Dierk VorwerkDVDierk Vorwerk
- 1 January 2004
- journal article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 230 (1) , 151-162
Abstract
To evaluate biologic response to poly(hydroxymethyl-p-xylylene-co-p-xylylene) (PHPX)-coated stents in vitro and in vivo in sheep. Physical stability, hemocompatibility, and cytotoxicity of the coating were first assessed in vitro. Thirty-six self-expanding nitinol (Memotherm), 24 stainless steel balloon-mounted (Palmaz), and 12 self-expanding nitinol (ZA) stents were coated with PHPX by using chemical vapor deposition polymerization. Seventy-two coated and 72 uncoated stents were placed into iliac arteries of 36 sheep. Sheep were classified into three groups of 12 animals each. In each group, six sheep were killed after 1 month; six, after 6 months. In each sheep, two uncoated stents were placed into one limb; two coated stents of the same type, into the opposite limb. In groups 1 and 2, Palmaz and Memotherm stents were used; in group 3, Memotherm and ZA stents were used. In groups 1 and 3, arteries were healthy. In group 2, arteries were pretreated with a Fogarty maneuver. Stent patency was measured with intravascular ultrasonography (US) and histologic analysis. Cellular response to coated and uncoated stents was assessed. Measurements were compared (Wilcoxon test). In vitro, PHPX coating was stable; hemocompatibility and cytotoxicity were similar to those of stainless steel. In vivo, patency of coated and uncoated Palmaz and ZA stents was not different (P >.05). Patency of coated and uncoated Memotherm stents did not differ in four of six follow-up subgroups, but it was significantly reduced in group 2 after 6 months (intravascular US, P =.03; histologic analysis, P =.01) and in group 3 after 1 month (histologic analysis, P =.01). Histologically, the cellular response to coated and uncoated stents was not different (P >.05). PHPX coating had good physical stability and biocompatibility in vitro and in vivo. Performance of coated and uncoated Palmaz and ZA stents was similar. Patency of Memotherm stents was similar in four of six follow-up subgroups. Materials effects did not result in severely enhanced neointimal hyperplasia.Keywords
This publication has 33 references indexed in Scilit:
- Comparison of the heparin coated vs the uncoated JostentS?no influence on restenosis or clinical outcomeEuropean Heart Journal, 2001
- Stent Coating With Titanium-Nitride-Oxide for Reduction of Neointimal HyperplasiaCirculation, 2001
- Sustained Local Delivery of Dexamethasone by a Novel Intravascular Eluting Stent to Prevent Restenosis in the Porcine Coronary Injury ModelJACC, 1997
- Marked Inflammatory Sequelae to Implantation of Biodegradable and Nonbiodegradable Polymers in Porcine Coronary ArteriesCirculation, 1996
- Biocompatibility of polymer-coated oversized metallic stents implanted in normal porcine coronary arteriesAtherosclerosis, 1995
- Intravascular stents: tissue-stent interactions and design considerations.American Journal of Roentgenology, 1993
- Perkutan implantierbare, durch Ballon aufdehnbare Gefäßprothese: Erste klinische ErgebnisseDMW - Deutsche Medizinische Wochenschrift, 1988
- Quantitation of 5-bromo-2-deoxyuridine incorporation into DNA: an enzyme immunoassay for the assessment of the lymphoid cell proliferative responseJournal of Immunological Methods, 1985
- Wetting: statics and dynamicsReviews of Modern Physics, 1985
- Growth of cultured calf aortic smooth muscle cells on cardiovascular prosthetic materialsJournal of Biomedical Materials Research, 1976