Dealing with Cerenkov radiation generated in organic scintillator dosimeters by bremsstrahlung beams
Authors:
- 30 March 2000
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 45 (5) , 1165-1182
Abstract
An organic scintillator detector system has been developed for radiotherapy bremsstrahlung dosimetry. The scintillators are connected to photodiodes by light pipes as the photodiodes must be removed and shielded from the incident radiation. The photodiodes see visible and near-visible light emissions from the scintillator as well as Cerenkov and fluorescence radiation that has been generated and trapped in the scintillator and light pipe. The Cerenkov and fluorescence radiation limits the accuracy of the dosimeter. This work examines a range of methods for diminishing the signal contribution of Cerenkov and fluorescence radiation while optimizing the scintillator signal. Three methods of achieving these goals have been used. They are: reflective coatings on the scintillator, long-wavelength-emitting scintillators used in conjunction with the photodiode, and absorptive filters placed between the light guide and photodiode. The contribution of the Cerenkov radiation to the light seen by the photodiode has been modelled and the model predictions have been tested using bremsstrahlung beams of peak energy between 13 and 20 MV, showing agreement with measurement.Keywords
This publication has 11 references indexed in Scilit:
- A new scintillator detector system for the quality assurance of60Co and high-energy therapy machinesPhysics in Medicine & Biology, 1994
- Optical filtering and spectral measurements of radiation-induced light in plastic scintillation dosimetryPhysics in Medicine & Biology, 1993
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical considerationsPhysics in Medicine & Biology, 1992
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurementsPhysics in Medicine & Biology, 1992
- Cerenkov light generated in optical fibres and other light pipes irradiated by electron beamsPhysics in Medicine & Biology, 1992
- Scintillating fibresNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1988
- Effects of radiation damage on p-type silicon detectorsPhysics in Medicine & Biology, 1983
- Measurement of the response of several organic scintillators to electrons, protons and deuteronsNuclear Instruments and Methods, 1968
- Response of Anthracene and Stilbene to Low-Energy Protons and X-RaysPhysical Review B, 1955
- Response of Anthracene Scintillation Crystals to Monoenergetic Soft X-RaysPhysical Review B, 1954