Purpose: To describe a model‐based reconstruction strategy for routine magnetic resonance imaging that accounts for gradient nonlinearity (GNL) during rather than after transformation to the image domain, and demonstrate that this approach reduces the spatial resolution loss that occurs during strictly image‐domain GNL‐correction.Methods: After reviewing conventional GNL‐correction methods, we propose a generic signal model for GNL‐affected magnetic resonance imaging acquisitions, discuss how it incorporates into contemporary image reconstruction platforms, and describe efficient nonuniform fast Fourier transform‐based computational routines for these. The impact of GNL‐correction on spatial resolution by the conventional and proposed approaches is investigated on phantom data acquired at varying offsets from gradient isocenter, as well as on fully sampled and (retrospectively) undersampled in vivo acquisitions.Results: Phantom results demonstrate that resolution loss that occurs during GNL‐correction is significantly less for the proposed strategy than for the standard approach at distances >10 cm from isocenter with a 35 cm field‐of‐view gradient coil. The in vivo results suggest that the proposed strategy better preserves fine anatomical detail than retrospective GNL‐correction while offering comparable geometric correction.Conclusion: Accounting for GNL during image reconstruction allows geometric distortion to be corrected with less spatial resolution loss than is typically observed with the conventional image domain correction strategy. Magn Reson Med 74:1019–1031, 2015.Keywords
Funding Information
- NIH (R01EB010065, R21EB017840)
This publication has 28 references indexed in Scilit:
- Sparsity and low‐contrast object detectabilityMagnetic Resonance in Medicine, 2011
- O‐space imaging: Highly efficient parallel imaging using second‐order nonlinear fields as encoding gradients with no phase encodingMagnetic Resonance in Medicine, 2010
- Parallel imaging in non-bijective, curvilinear magnetic field gradients: a concept studyMagnetic Resonance Materials in Physics, Biology and Medicine, 2008
- Investigation of MR image distortion for radiotherapy treatment planning of prostate cancerPhysics in Medicine & Biology, 2006
- Iterative decomposition of water and fat with echo asymmetry and least‐squares estimation (IDEAL): Application with fast spin‐echo imagingMagnetic Resonance in Medicine, 2005
- A complete distortion correction for MR images: I. Gradient warp correctionPhysics in Medicine & Biology, 2005
- Correction for gradient nonlinearity in continuously moving table MR imagingMagnetic Resonance in Medicine, 2004
- Use of spherical harmonic deconvolution methods to compensate for nonlinear gradient effects on MRI imagesMagnetic Resonance in Medicine, 2004
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- Estimation of the Mean of a Multivariate Normal DistributionThe Annals of Statistics, 1981