Interference of medical contrast media on laboratory testing
Authors:
Open Access
- 1 January 2014
- journal article
- review article
- Published by Croatian Society for Medical Biochemistry and Laboratory Medicine in Biochemia Medica
- Vol. 24 (1) , 80-8
Abstract
The use of contrast media such as organic iodine molecules and gadolinium contrast agents is commonplace in diagnostic imaging. Although there is widespread perception that side effects and drug interactions may be the leading problems caused by these compounds, various degrees of interference with some laboratory tests have been clearly demonstrated. Overall, the described interference for iodinate contrast media include inappropriate gel barrier formation in blood tubes, the appearance of abnormal peaks in capillary zone electrophoresis of serum proteins, and a positive bias in assessment of cardiac troponin I with one immunoassay. The interference for gadolinium contrast agents include negative bias in calcium assessment with ortho-cresolphthalein colorimetric assays and occasional positive bias using some Arsenazo reagents, negative bias in measurement of angiotensin converting enzyme (ACE) and zinc (colorimetric assay), as well as positive bias in creatinine (Jaffe reaction), total iron binding capacity (TIBC, ferrozine method), magnesium (calmagite reagent) and selenium (mass spectrometry) measurement. Interference has also been reported in assessment of serum indices, pulse oximetry and methaemoglobin in samples of patients receiving Patent Blue V. Under several circumstances the interference was absent from manufacturer-supplied information and limited to certain type of reagents and/or analytes, so that local verification may be advisable to establish whether or not the test in use may be biased. Since the elimination half-life of these compounds is typically lower than 2 h, blood collection after this period may be a safer alternative in patients who have received contrast media for diagnostic purposes.Keywords
This publication has 52 references indexed in Scilit:
- Effect of Patent Blue V ® dye on sample interference indices on the Abbott ArchitectAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2012
- An Unusual Case of a Primary Blood Collection Tube with Floating Separator GelJournal of Clinical Laboratory Analysis, 2012
- Preanalytical phase – a continuous challenge for laboratory professionalsBiochemia Medica, 2012
- Unusual false‐positive case of urinary screening for buprenorphineJournal of Clinical Laboratory Analysis, 2011
- Interference testing.2008
- Clinical and Economic Impact of Falsely Decreased Calcium Values Caused by Gadoversetamide InterferenceAmerican Journal of Roentgenology, 2008
- Measurement of Serum Calcium Concentration After Administration of Four Gadolinium-Based Contrast Agents to Human VolunteersAmerican Journal of Roentgenology, 2007
- Pseudoparaproteinemia after Iopamidol Infusion for Coronary AngiographyClinical Chemistry, 2005
- Iodinated Contrast Media Interfere with Gel Barrier Formation in Plasma and Serum Separator TubesClinical Chemistry, 2003
- Interference in ImmunoassayAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1999