Validation of a Miniaturized Test Loop for the Assessment of Human Blood Damage by Continuous-Flow Left-Ventricular Assist Devices
Open Access
- 24 August 2021
- journal article
- research article
- Published by Springer Nature in Annals of Biomedical Engineering
- Vol. 49 (12) , 3165-3175
Abstract
Despite improved hemocompatibility of left-ventricular assist devices (LVADs), assessment of blood damage remains mandatory in preclinical testing as standardized by ASTM-F1841. The most relevant test fluid is fresh, non-pooled human blood, but the limited volume of a standard donation requires significantly smaller loops than those commonly used with animal blood. In a recent study with porcine blood, we verified a miniaturized test loop with only 160 mL for the ASTM-conform paired testing of at least two LVADs and a static reference. Here, we validated this mini test loop for standardized assessment of blood damage with one 450-mL single donation of fresh human blood. Blood damage was assessed for HeartMate 3 and BPX-80 in 9 experiments with heparinized human blood for 6 hours. We analyzed plasma free hemoglobin, von Willebrand factor (vWF) concentration and collagen-binding functionality and calculated indices of hemolysis and vWF-ratios. Overall, we observed less blood damage compared to our previous study; however, the differences in mean indices of hemolysis and in mean normalized vWF-ratio between BPX-80 and HeartMate 3 were consistent for human blood. Thus, our mini test loop proved to be valid for preclinical standardized assessment of blood damage with only 450 mL of fresh human blood.Keywords
Funding Information
- RWTH Aachen University
This publication has 30 references indexed in Scilit:
- A Reusable, Compliant, Small Volume Blood Reservoir for In Vitro Hemolysis TestingArtificial Organs, 2016
- Mechanical Blood Trauma in Assisted Circulation: Sublethal RBC Damage Preceding HemolysisThe International Journal of Artificial Organs, 2016
- Circulatory support devices: fundamental aspects and clinical management of bleeding and thrombosisJournal of Thrombosis and Haemostasis, 2015
- Numerical Analysis of Blood Damage Potential of the HeartMate II and HeartWare HVAD Rotary Blood PumpsArtificial Organs, 2015
- Shear-Induced Hemolysis: Species DifferencesArtificial Organs, 2015
- Platelet glycoprotein Ibα ectodomain shedding and non-surgical bleeding in heart failure patients supported by continuous-flow left ventricular assist devicesThe Journal of Heart and Lung Transplantation, 2014
- Axial and centrifugal continuous-flow rotary pumps: A translation from pump mechanics to clinical practiceThe Journal of Heart and Lung Transplantation, 2013
- Heparin promotes platelet responsiveness by potentiating αIIbβ3-mediated outside-in signalingBlood, 2011
- Diagnostic workup of patients with acquired von Willebrand syndrome: a retrospective single‐centre cohort studyJournal of Thrombosis and Haemostasis, 2008
- Detecting outliers when fitting data with nonlinear regression – a new method based on robust nonlinear regression and the false discovery rateBMC Bioinformatics, 2006