Instrument-Free Control of the Standard Potential of Potentiometric Solid-Contact Ion-Selective Electrodes by Short-Circuiting with a Conventional Reference Electrode
Authors:
- 4 October 2014
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 86 (21) , 10540-10545
Abstract
A simple, instrument-free method to control the standard potential (E°) of potentiometric solid-contact ion-selective electrodes (SC-ISE) is described. In this method, the electrode potential of a SC-ISE is reset by short-circuiting the electrode with a metallic wire to a conventional Ag/AgCl/3 M KCl reference electrode (RE) in a solution containing primary ions. The method is studied experimentally for SC-ISEs where the conducting polymer poly(3,4-ethylenedioxythiophene) doped with the bulky anion poly(sodium 4-styrenesulfonate), PEDOT(PSS), is used as the solid contact. Three different types of ion-selective membranes (ISMs) are studied: two potassium-selective membranes, with and without the lipohilic additive tetradodecylammonium tetrakis(4-chlorophenyl)borate (ETH-500) and a cation-sensitive membrane without an ionophore. When the SC-ISE is short-circuited with the RE, the PEDOT(PSS) layer is oxidized or reduced, thereby shifting the potential of the SC-ISE to the proximity of the potential of the RE so that the potential difference between these two electrodes becomes zero or close to zero. The slope of the calibration curve is preserved after the short-circuit treatment of the SC-ISEs. The short-circuiting method is an important step toward calibration-free potentiometric analysis.Keywords
This publication has 34 references indexed in Scilit:
- A solid electrolyte sensor for trace gas analysisSensors and Actuators B: Chemical, 2013
- Background current elimination in thin layer ion-selective membrane coulometryElectrochemistry Communications, 2010
- Diagnostic of functionality of polymer membrane – based ion selective electrodes by impedance spectroscopyAnalytical Methods, 2010
- Thin Layer Coulometry with Ionophore Based Ion-Selective MembranesAnalytical Chemistry, 2010
- Quality control criteria for solid-contact, solvent polymeric membrane ion-selective electrodesJournal of Solid State Electrochemistry, 2008
- Beyond potentiometry: Robust electrochemical ion sensor concepts in view of remote chemical sensingTalanta, 2008
- Modern directions for potentiometric sensorsJournal of the Brazilian Chemical Society, 2008
- Highly Selective All‐Plastic, Disposable, Cu2+‐Selective ElectrodesElectroanalysis, 2005
- A New Calibration Free pH‐Probe for In Situ Measurements of Soil pHElectroanalysis, 2004
- Peer Reviewed: Internet-Scale Sensing2004