Optimized Scanning Conditions of High Resolution CT in the Follow-Up of Pulmonary Emphysema

Authors:
  • HIHarumi Itoh
    Harumi Itoh
  • HSHiroaki Sakai
    Hiroaki Sakai
  • YNYasutaka Nakano
    Yasutaka Nakano
  • SMShigeo Muro
    Shigeo Muro
  • THToyohiro Hirai
    Toyohiro Hirai
  • YTYasutaka Takubo
    Yasutaka Takubo
  • KCKazuo Chin
    Kazuo Chin
  • MOMotoharu Ohi
    Motoharu Ohi
  • KNKoichi Nishimura
    Koichi Nishimura
  • KYKazuhiro Yamaguchi
    Kazuhiro Yamaguchi
  • TNTakashi Nakamura
    Takashi Nakamura

Abstract
To address the optimal scanning condition of high resolution computerized tomography from the perspective of minimizing exposed dose. The influence of the electric current, the slice number, and the slice thickness on precise percent ratio of the low attenuation area to whole lung field (LAA%) of chronic obstructive pulmonary disease patients was examined. The standard conditions were 250 mA, 3 slices, 2 mm slice thickness, and a varied parameter. In cases showing an LAA% less than 30, LAA% obtained by ≤ 150 mA were significantly larger than those by 250 mA. The mean LAA% with 3 and 10 slices were well correlated and the correlation with lung function was similar. The correlation of LAA% with lung function was approximately the same between the 2 and 5 mm slice thickness. The electric current must be ≥200 mA, and 3 slices and 2 mm slice thickness are appropriate.