Validity and repeatability of the Aladdin ocular biometer
Authors:
- Emma J BerrowEBEmma J Berrow
- Shehzad A NarooSNShehzad A Naroo
- James S WolffsohnJames S Wolffsohn
- Detlef UthoffDUDetlef Uthoff
- Detlef HollandDHDetlef Holland
- Sunil ShahSSSunil Shah
- 13 November 2013
- journal article
- research article
- Published by BMJ in British Journal of Ophthalmology
- Vol. 98 (2) , 256-258
Abstract
To assess the accuracy and reproducibility of biometry undertaken with the Aladdin (Topcon, Tokyo, Japan) in comparison with the current gold standard device, the IOLMaster 500 (Zeiss, Jena, Germany). University Eye Clinic, Birmingham, UK and Refractive Surgery Centre, Kiel, Germany. The right eye of 75 patients with cataracts and 22 healthy participants were assessed using the two devices. Measurements of axial length (AL), anterior chamber depth (ACD) and keratometry (K) were undertaken with the Aladdin and IOLMaster 500 in random order by an experienced practitioner. A second practitioner then obtained measurements for each participant using the Aladdin biometer in order to assess interobserver variability. No statistically significant differences (p>0.05) between the two biometers were found for average difference (AL)±95% CI=0.01±0.06 mm), ACD (0.00±0.11 mm) or mean K values (0.08±0.51 D). Furthermore, interobserver variability was very good for each parameter (weighted κ≥0.85). One patient's IOL powers could not be calculated with either biometer measurements, whereas a further three could not be analysed by the IOLMaster 500. The IOL power calculated from the valid measurements was not statistically significantly different between the biometers (p=0.842), with 91% of predictions within±0.25 D. The Aladdin is a quick, easy-to-use biometer that produces valid and reproducible results that are comparable with those obtained with the IOLMaster 500.Keywords
This publication has 15 references indexed in Scilit:
- Assessment of a new averaging algorithm to increase the sensitivity of axial eye length measurement with optical biometry in eyes with dense cataractJournal of Cataract & Refractive Surgery, 2011
- Evaluation of a new IOLMaster algorithm to measure axial lengthJournal of Cataract & Refractive Surgery, 2008
- Sources of error in intraocular lens power calculationJournal of Cataract & Refractive Surgery, 2008
- Biometry and intraocular lens power calculationCurrent Opinion in Ophthalmology, 2008
- Calculation of intraocular lens power: a reviewActa Ophthalmologica Scandinavica, 2007
- Reliability of the IOLMaster in axial length evaluation in silicone oil-filled eyesEye, 2006
- Evaluation of the practicality of optical biometry and applanation ultrasound in 253 eyesJournal of Cataract & Refractive Surgery, 2003
- Immersion A-scan compared with partial coherence interferometryJournal of Cataract & Refractive Surgery, 2002
- Partial coherence interferometry: a novel approach to biometry in cataract surgeryAmerican Journal of Ophthalmology, 1998
- Comparison of contact and immersion techniques for axial length measurement and implant power calculationJournal of Cataract & Refractive Surgery, 1989