Brain PET motion correction using 3D face-shape model: the first clinical study
Open Access
- 19 July 2022
- journal article
- research article
- Published by Springer Nature in Annals of Nuclear Medicine
- Vol. 36 (10) , 904-912
Abstract
Objective: Head motions during brain PET scan cause degradation of brain images, but head fixation or external-maker attachment become burdensome on patients. Therefore, we have developed a motion correction method that uses a 3D face-shape model generated by a range-sensing camera (Kinect) and by CT images. We have successfully corrected the PET images of a moving mannequin-head phantom containing radioactivity. Here, we conducted a volunteer study to verify the effectiveness of our method for clinical data. Methods: Eight healthy men volunteers aged 22–45 years underwent a 10-min head-fixed PET scan as a standard of truth in this study, which was started 45 min after 18F-fluorodeoxyglucose (285 ± 23 MBq) injection, and followed by a 15-min head-moving PET scan with the developed Kinect based motion-tracking system. First, selecting a motion-less period of the head-moving PET scan provided a reference PET image. Second, CT images separately obtained on the same day were registered to the reference PET image, and create a 3D face-shape model, then, to which Kinect-based 3D face-shape model matched. This matching parameter was used for spatial calibration between the Kinect and the PET system. This calibration parameter and the motion-tracking of the 3D face shape by Kinect comprised our motion correction method. The head-moving PET with motion correction was compared with the head-fixed PET images visually and by standard uptake value ratios (SUVRs) in the seven volume-of-interest regions. To confirm the spatial calibration accuracy, a test–retest experiment was performed by repeating the head-moving PET with motion correction twice where the volunteer’s pose and the sensor’s position were different. Results: No difference was identified visually and statistically in SUVRs between the head-moving PET images with motion correction and the head-fixed PET images. One of the small nuclei, the inferior colliculus, was identified in the head-fixed PET images and in the head-moving PET images with motion correction, but not in those without motion correction. In the test–retest experiment, the SUVRs were well correlated (determinant coefficient, r2 = 0.995). Conclusion: Our motion correction method provided good accuracy for the volunteer data which suggested it is useable in clinical settings.Keywords
Funding Information
- Japan Society for the Promotion of Science (17K18376)
This publication has 25 references indexed in Scilit:
- Repurposing the Microsoft Kinect for Windows v2 for external head motion tracking for brain PETPhysics in Medicine & Biology, 2015
- Florbetaben PET imaging to detect amyloid beta plaques in Alzheimer's disease: Phase 3 studyAlzheimer's & Dementia, 2015
- Multicentre population‐based dementia prevalence survey in Japan: a preliminary reportPsychogeriatrics, 2012
- Use of Florbetapir-PET for Imaging β-Amyloid PathologyJAMA, 2011
- Motion Tracking in Narrow Spaces: A Structured Light ApproachLecture Notes in Computer Science, 2010
- Improvement of the spatial resolution of the MicroPET R4 scanner by wobbling the bedMedical Physics, 2008
- New, faster, image-based scatter correction for 3D PETIEEE Transactions on Nuclear Science, 2000
- A method for registration of 3-D shapesIEEE Transactions on Pattern Analysis and Machine Intelligence, 1992
- The human auditory brain stem as a generator of auditory evoked potentialsHearing Research, 1987
- A Simplex Method for Function MinimizationThe Computer Journal, 1965