Monte Carlo simulation of the Tomotherapy treatment unit in the static mode using MC HAMMER, a Monte Carlo tool dedicated to Tomotherapy
- 1 June 2007
- conference paper
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 74 (1) , 021019
Abstract
Helical tomotherapy (HT) is designed to deliver highly modulated IMRT treatments. The concept of HT provides new challenges in MC simulation, because simultaneous movement of the gantry, the couch and the multi-leaf collimator (MLC) must be simulated accurately. However, before accounting for gantry, couch movement and multileaf collimator configurations, high accuracy must be achieved while simulating open static fields (1 × 40, 2.5 × 40 and 5 × 40 cm2). This is performed using MC HAMMER, which is a graphical user interface allowing MC simulation using PENELOPE for various configurations of HT. Since the geometry of the different elements and materials involved in the beam generation are precisely known and defined, the only parameters that need to be tuned on are therefore electron source spot size and electron energy. Beyond the build up region, good agreement (2%/1mm) is achieved for all the field sizes between measurements (ion chamber) and simulations with an electron source energy set to 5.5 MeV. The electron source spot size is modelled as a gaussian distribution with full width half maximum equal to 1.4 mm. This value was chosen to match measured and calculated penumbras in the longitudinal direction.Keywords
This publication has 20 references indexed in Scilit:
- Parotid-sparing intensity-modulated radiotherapy (IMRT) for nasopharyngeal carcinoma: Preserved parotid function after IMRT on quantitative salivary scintigraphy, and comparison with historical data after conventional radiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- Comparison of inhomogeneity correction algorithms in small photon fieldsMedical Physics, 2005
- Accurate convolution/superposition for multi-resolution dose calculation using cumulative tabulated kernelsPhysics in Medicine & Biology, 2005
- Comparison of dose calculation algorithms in phantoms with lung equivalent heterogeneities under conditions of lateral electronic disequilibriumMedical Physics, 2004
- Dosimetry and radiobiologic model comparison of IMRT and 3D conformal radiotherapy in treatment of carcinoma of the prostateInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2004
- Experimental verification of convolution/superposition photon dose calculations for radiotherapy treatment planningPhysics in Medicine & Biology, 2003
- Underdosage of the upper‐airway mucosa for small fields as used in intensity‐modulated radiation therapy: A comparison between radiochromic film measurements, Monte Carlo simulations, and collapsed cone convolution calculationsMedical Physics, 2002
- Three-dimensional intensity-modulated radiotherapy in the treatment of nasopharyngeal carcinoma: the University of California–San Francisco experienceInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2000
- The impact of electron transport on the accuracy of computed doseMedical Physics, 2000
- Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous mediaMedical Physics, 1989