Verification of an Electromagnetic Placement Device Compared With Abdominal Radiograph to Predict Accuracy of Feeding Tube Placement
Authors:
- 23 June 2011
- journal article
- research article
- Published by Wiley in Journal of Parenteral and Enteral Nutrition
- Vol. 35 (4) , 535-539
Abstract
Background: Use of an electromagnetic placement device (EMPD) facilitates placement of feeding tubes at the bedside. Standard practice for verification of feeding tube placement is via radiographic confirmation. The purpose of this research study was to assess the accuracy of placement of small‐bore feeding tubes (SBFTs) as determined by EMPD interpretation compared with that of abdominal radiograph verification by a radiologist. Methods: This multicenter prospective study enrolled patients requiring bedside feeding tube placement. SBFTs were placed by an experienced investigator using the EMPD. Two abdominal radiographs were then obtained: one after initial SBFT placement and an additional radiograph after injection of contrast. Documentation of location based on clinician interpretation using the EMPD was then compared with radiologist interpretation. Results: The final sample size was 194 patients, including 18 pediatric patients. Patient age ranged from 12 days to 102 years. Median time for tube placement was 12 minutes. Of the 194 patients, only 1 patient had data showing discrepancies between the original EMPD verification and the final abdominal radiograph interpretation, providing a 99.5% agreement. No patient experienced complications during SBFT placement, and 15 patients had inadvertent airway placement that was avoided with the use of the EMPD. Conclusions: There was a high percentage of agreement between EMPD and radiologic interpretation after contrast injection. The EMPD aided in avoiding inadvertent airway placement, with no patient complications. This device can be used safely at the bedside to facilitate placement of feeding tubes, leading to the delivery of early enteral nutrition.Keywords
This publication has 26 references indexed in Scilit:
- Bedside Electromagnetic‐Guided Feeding Tube Placement: An Improvement Over Traditional Placement Technique?Nutrition in Clinical Practice, 2007
- Complications related to feeding tube placementCurrent Opinion in Gastroenterology, 2007
- Endoscopic placement of enteral feeding cathetersCurrent Opinion in Gastroenterology, 2006
- Enhancing Patient Safety During Feeding‐Tube Insertion: A Review of More Than 2000 InsertionsJournal of Parenteral and Enteral Nutrition, 2006
- Technologic Approaches to Determining Proper Placement of Enteral Feeding TubesAACN Advanced Critical Care, 2006
- Placement of nasoenteral feeding tubes using external magnetic guidanceJournal of Parenteral and Enteral Nutrition, 2004
- Air insufflation technique of enteral tube insertionCritical Care Medicine, 1998
- Fluoroscopically Guided Nasoenteric Feeding Tube Placement Versus Bedside PlacementSouthern Medical Journal, 1995
- Bedside Enteral Feeding Tube Placement into Duodenum and JejunumJournal of Parenteral and Enteral Nutrition, 1986
- Accidental intrapulmonary ClinifeedAnaesthesia, 1981