Lactate Measurement Interference by Hemoglobin-Based Oxygen Carriers (Oxyglobin??, Hemopure??, and Hemolink???)
Authors:
- Jonathan S. JahrJJJonathan S. Jahr
- Stephen OsgoodSOStephen Osgood
- Stephen J. RothenbergSRStephen J. Rothenberg
- Qiao-Ling LiQLQiao-Ling Li
- Anthony W. ButchABAnthony W. Butch
- Robert GuntherRGRobert Gunther
- Anthony CheungACAnthony Cheung
- Bernd DriessenBDBernd Driessen
- 1 February 2005
- journal article
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 100 (2) , 431-436
Abstract
Trated L-lactate solution, HBOC, and blood or plasma with added PlasmaLyte-A™ were added to sample tubes to make a linear and constant increase in lactate concentration in consecutive samples. Sample lactate concentrations ranged from 5–110 mg/dL (0.6–12 mm) (physiological reference range: 5–20 mg/dL [0.56–2.2 mm]). Comparisons were made between machine measured lactate concentrations and calculated lactate concentrations. For Hb glutamer-250, the average difference between measured and calculated lactate concentrations was −5.1 mg/dL (−0.57 mm) (LX-20®), with greater underestimation at larger lactate concentrations. For Hb raffimer, the average difference was −2.2 mg/dL (−0.24 mm) (LX-20®). The veterinary product, Hb glutamer-200, was tested on 3 analyzers (LX-20®, YSI 1500, and YSI 2300). The YSI 1500 was the most accurate instrument with the mean difference between measured minus calculated lactate being +1.3 mg/dL versus −2.6 mg/dL (YSI 2300) and −8.4 mg/dL (LX-20®). The clinical implications of this study are that with increasing levels of an HBOC in plasma, lactate interpretation may become inaccurate, especially at larger lactate concentrations, causing underestimation of measured lactate values and possible under-treatment of the patient. Therefore, caution must be exercised when interpreting lactate results when a HBOC is present in plasma....Keywords
This publication has 20 references indexed in Scilit:
- Arterial oxygenation and oxygen delivery after hemoglobin-based oxygen carrier infusion in canine hypovolemic shock: A dose-response studyCritical Care Medicine, 2003
- Current status of blood substitute research: towards a new paradigmJournal of Internal Medicine, 2003
- Redox side reactions of haemoglobin and cell signalling mechanisms1Journal of Internal Medicine, 2003
- Hemoglobin-based oxygen carriersCurrent Opinion in Hematology, 2002
- Effects of Hemoglobin Glutamer-250 (Bovine) (HBOC-201, Hemopure) on Coagulation TestingAmerican Journal of Therapeutics, 2002
- Blood Substitutes and Oxygen Therapeutics: An Overview and Current StatusAmerican Journal of Therapeutics, 2002
- Haemoglobin-based erythrocyte transfusion substitutesExpert Opinion on Biological Therapy, 2001
- Multiple Regression Analysis of Interference Effects from a Hemoglobin-Based Oxygen Carrier SolutionClinical Chemistry and Laboratory Medicine (CCLM), 1999
- Effect of the Hemoglobin-Based Oxygen Carrier HBOC-201 on Laboratory Instrumentation: Cobas Integra, Chiron Blood Gas Analyzer 840, Sysmex™SE-9000 and BCTClinical Chemistry and Laboratory Medicine (CCLM), 1999
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986