Enhanced detection of electrode placement/connection errors
Authors:
- 1 September 2008
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 02766574,p. 89-92
Abstract
Lead connection and electrode positional errors are a common problem in ECG recording. This study set out to review the sensitivity and specificity of existing criteria in the Glasgow program using an older (1997) version of the software and to produce enhancements where required for incorporation into the current version of the program still in development. 50 volunteers were recruited to the study. Arm and leg lead connection errors were introduced as were V1/V2 and V2/V3 connection reversals. It was shown that detection of arm lead connection errors could be enhanced from 64% to 88% at 100% specificity. Chest lead misconnections were detected with improved sensitivity. V1 and V2 reversal was much more easily detected than V2 and V3 reversal while maintaining high specificity.Keywords
This publication has 6 references indexed in Scilit:
- Errors in the computerized electrocardiogram interpretation of cardiac rhythmJournal of Electrocardiology, 2007
- The university of glasgow (Uni-G) ECG analysis program2005
- Detection of frequently overlooked electrocardiographic lead reversals using artificial neural networksThe American Journal of Cardiology, 1996
- Artificial neural networks for recognition of electrocardiographic lead reversalThe American Journal of Cardiology, 1995
- The Diagnostic Performance of Computer Programs for the Interpretation of ElectrocardiogramsNew England Journal of Medicine, 1991
- A reference data base for multilead electrocardiographic computer measurement programsJACC, 1987