The DosiMap, a new 2D scintillating dosimeter for IMRT quality assurance: Characterization of two Čerenkov discrimination methods
Authors:
- A‐M. FrelinAFA‐M. Frelin
- J‐M. FontbonneJFJ‐M. Fontbonne
- G. BanGBG. Ban
- J. ColinJCJ. Colin
- M. LabalmeMLM. Labalme
- A. BatallaABA. Batalla
- A. VelaAVA. Vela
- P. BoherPBP. Boher
- M. BraudMBM. Braud
- T. LerouxTLT. Leroux
- 7 April 2008
- journal article
- Published by Wiley in Medical Physics
- Vol. 35 (5) , 1651-1662
Abstract
New radiation therapy techniques such as IMRT present significant efficiency due to their highly conformal dose distributions. A consequence of the complexity of their dose distributions (high gradients, small irradiation fields, low dose distribution, …) is the requirement for better precision quality assurance than in classical radiotherapy in order to compare the conformation of the delivered dose with the planned dose distribution and to guarantee the quality of the treatment. Currently this control is mostly performed by matrices of ionization chambers, diode detectors, dosimetric films, portal imaging, or dosimetric gels. Another approach is scintillation dosimetry, which has been developed in the last 15 years mainly through scintillating fiber devices. Despite having many advantages over other methods it is still at an experimental level for routine dosimetry because the Čerenkov radiation produced under irradiation represents an important stem effect. A new 2D water equivalent scintillating dosimeter, the DosiMap, and two different Čerenkov discrimination methods were developed with the collaboration of the Laboratoire de Physique Corpusculaire of Caen, the Comprehensive Cancer Center François Baclesse, and the ELDIM Co., in the frame of the MAESTRO European project. The DosiMap consists of a plastic scintillating sheet placed inside a transparent polystyrene phantom. The light distribution produced under irradiation is recorded by a CCD camera. Our first Čerenkov discrimination technique is subtractive. It uses a chessboard pattern placed in front of the scintillator, which provides a background signal containing only Čerenkov light. Our second discrimination technique is colorimetric. It performs a spectral analysis of the light signal, which allows the unfolding of the Čerenkov radiation and the scintillation. Tests were carried out with our DosiMap prototype and the performances of the two discrimination methods were assessed. The comparison of the dose measurements performed with the DosiMap and with dosimetric films for three different irradiation configurations showed discrepancies smaller than for a 2 mm spatial resolution. Two innovative discrimination solutions were demonstrated to separate the scintillation from the Čerenkov radiation. It was also shown that the DosiMap, which is water equivalent, fast, and user friendly, is a very promising tool for radiotherapy quality assurance.Keywords
This publication has 25 references indexed in Scilit:
- Analysis of the sources of uncertainty for EDR2 film‐based IMRT quality assuranceJournal of Applied Clinical Medical Physics, 2006
- A technical evaluation of the Nucletron FIRST system: Conformance of a remote afterloading brachytherapy seed implantation system to manufacturer specifications and AAPM Task Group report recommendationsJournal of Applied Clinical Medical Physics, 2005
- Scintillating fiber dosimeter for radiation therapy acceleratorIEEE Transactions on Nuclear Science, 2002
- A miniature "scintillator-fiberoptic-PMT" detector system for the dosimetry of small fields in stereotactic radiosurgeryIEEE Transactions on Nuclear Science, 2001
- Performance of a fluorescent screen and CCD camera as a two-dimensional dosimetry system for dynamic treatment techniquesMedical Physics, 2000
- Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapyMedical Physics, 1998
- Optical filtering and spectral measurements of radiation-induced light in plastic scintillation dosimetryPhysics in Medicine & Biology, 1993
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical considerationsPhysics in Medicine & Biology, 1992
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurementsPhysics in Medicine & Biology, 1992
- Cerenkov light generated in optical fibres and other light pipes irradiated by electron beamsPhysics in Medicine & Biology, 1992