Extrusion Rates and Hearing Results in Ossicular Reconstruction

Authors:

Abstract
OBJECTIVES: To examine hearing results, extrusion rates, and factors likely to affect outcome in patients who underwent ossicular chain reconstruction with Plasti‐Pore and hydroxylapatite total ossicular replacement prostheses (TORPs) and partial ossicular replacement prosthesis (PORPs).STUDY DESIGN AND SETTING: Retrospective chart review was used. Subjects were 1210 consecutive tympanoplasties with or without mastoidectomy that involved ossicular reconstruction with TORPs (n = 560) or PORPs (n = 650) performed in a tertiary referral neurotologic private practice. Only 20% of the cases were primary surgeries, with the majority planned second stage or revision procedures.RESULTS: Average last postoperative air‐bone gap (ABG) was 19.2 dB with closure of the ABG to within 20 dB in 62.9%. Hearing results were better for cases who had not had previous surgery, in those with a diagnosis other than chronic otitis media, when a cartilage graft was used, and for Plasti‐Pore rather than hydroxylapatite. Extrusion was known to occur in 4%.CONCLUSION: After more than 25 years of use in our clinic, Plasti‐Pore ossicular replacement pros‐theses continue to provide reliable hearing results with a low rate of extrusion.