Monitoring of end tidal carbon dioxide and transcutaneous carbon dioxide during neonatal transport
Open Access
- 29 April 2005
- journal article
- research article
- Published by BMJ in Archives of Disease in Childhood - Fetal and Neonatal Edition
- Vol. 90 (6) , F523-F526
Abstract
Objective: To assess the accuracy of measurements of end tidal carbon dioxide (CO2) during neonatal transport compared with arterial and transcutaneous measurements. Design: Paired end tidal and transcutaneous CO2 recordings were taken frequently during road transport of 21 ventilated neonates. The first paired CO2 values were compared with an arterial blood gas. The differences between arterial CO2 (Paco 2), transcutaneous CO2 (TcPco 2), and end tidal CO2 (Petco 2) were analysed. The Bland-Altman method was used to assess bias and repeatability. Results: Petco 2 correlated strongly with Paco 2 and TcPco 2. However, Petco 2 underestimated Paco 2 at a clinically unacceptable level (mean (SD) 1.1 (0.70) kPa) and did not trend reliably over time within individual subjects. The Petco 2 bias was independent of Paco 2 and severity of lung disease. Conclusions: Petco 2 had an unacceptable under-recording bias. TcPco 2 should currently be considered the preferred method of non-invasive CO2 monitoring for neonatal transport.Keywords
This publication has 20 references indexed in Scilit:
- Transcutaneous measurement of carbon dioxide tension during long-distance transport of neonates receiving mechanical ventilation.1998
- Mainstream End-tidal Carbon Dioxide Monitoring in the Neonatal Intensive Care UnitPediatrics, 1998
- Evaluation of a Portable Infrared End-Tidal Carbon Dioxide Monitor During Pediatric Interhospital TransportPediatrics, 1995
- Capnometry and anaesthesiaCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1992
- Estimation of arterial carbon dioxide by end‐tidal and transcutaneous P measurements in ventilated childrenPediatric Pulmonology, 1992
- End-tidal carbon dioxide pressure in neonates and infants measured by aspiration and flow-through capnographyJournal of Clinical Monitoring and Computing, 1991
- Continuous end-tidal CO2 sampling within the proximal endotracheal tube estimates arterial CO2 tension in infantsCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1991
- Is Distal Sampling of End-tidal CO2Necessary in Small Subjects?Anesthesiology, 1990
- Accuracy of end-tidal PCO2 measurements using a sidestream capnometer in infants and children ventilated with the Sechrist infant ventilatorCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1990
- Transcutaneous carbon dioxide monitoring during neonatal transportCritical Care Medicine, 1986