Abstract
Objective: To assess the accuracy of measurements of end tidal carbon dioxide (CO2) during neonatal transport compared with arterial and transcutaneous measurements. Design: Paired end tidal and transcutaneous CO2 recordings were taken frequently during road transport of 21 ventilated neonates. The first paired CO2 values were compared with an arterial blood gas. The differences between arterial CO2 (Paco 2), transcutaneous CO2 (TcPco 2), and end tidal CO2 (Petco 2) were analysed. The Bland-Altman method was used to assess bias and repeatability. Results: Petco 2 correlated strongly with Paco 2 and TcPco 2. However, Petco 2 underestimated Paco 2 at a clinically unacceptable level (mean (SD) 1.1 (0.70) kPa) and did not trend reliably over time within individual subjects. The Petco 2 bias was independent of Paco 2 and severity of lung disease. Conclusions: Petco 2 had an unacceptable under-recording bias. TcPco 2 should currently be considered the preferred method of non-invasive CO2 monitoring for neonatal transport.