Nine Orders of Magnitude Dynamic Range: Picomolar to Millimolar Concentration Measurement in Capillary Electrophoresis with Laser Induced Fluorescence Detection Employing Cascaded Avalanche Photodiode Photon Counters
- 16 March 2011
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 83 (7) , 2748-2753
Abstract
The dynamic range of capillary electrophoresis analysis is ultimately limited by molecular shot noise at low concentrations and by concentration-induced band broadening at high concentrations. We report a system that approaches these fundamental limits. A laser-induced fluorescence detector is reported that employs a cascade of four fiber-optic beam splitters connected in series to generate a primary signal and four attenuated signals, each monitored by a single-photon counting avalanche photodiode. Appropriate scaling of the signals from the five photodiodes produces a linear optical calibration curve for 5-carboxyl-tetramethylrhodamine from the concentration detection limit of 1 pM to the upper limit of 1 mM. Mass detection limits are 120 yoctomoles (70 molecules) injected into the instrument. The very-wide dynamic range instrument was used to study the metabolic products of the fluorescently labeled glycosphingolipid tetramethylrhodamine labeled GM1 (GM1-TMR) produced by single cells isolated from the rat cerebellum.Keywords
This publication has 26 references indexed in Scilit:
- Statistical Treatment of Photon/Electron Counting: Extending the Linear Dynamic Range from the Dark Count Rate to SaturationAnalytical Chemistry, 2010
- Six orders of magnitude dynamic range in capillary electrophoresis with ultrasensitive laser-induced fluorescence detectionTalanta, 2009
- Femtomolar Concentration Detection Limit and Zeptomole Mass Detection Limit for Protein Separation by Capillary Isoelectric Focusing and Laser-Induced Fluorescence DetectionAnalytical Chemistry, 2009
- Gangliosides and Nogo Receptors Independently Mediate Myelin-associated Glycoprotein Inhibition of Neurite Outgrowth in Different Nerve CellsJournal of Biological Chemistry, 2007
- Yoctomole analysis of ganglioside metabolism in PC12 cellular homogenatesElectrophoresis, 2007
- Synthesis of reference standards to enable single cell metabolomic studies of tetramethylrhodamine-labeled ganglioside GM1Carbohydrate Research, 2007
- Asymmetry between Sister Cells in a Cancer Cell Line Revealed by Chemical CytometryAnalytical Chemistry, 2004
- Effect of detector nonlinearity on the height, area, width, and moments of peaks in liquid chromatography with absorbance detectorsAnalytical Chemistry, 1981
- Low absorbance spectrophotometryAnalytical Chemistry, 1981
- Efficiency of Counting SystemsProceedings of the Physical Society (1926-1948), 1948