Abstract
Multielectrode catheters using duty-cycled radiofrequency (RF) have been developed to treat atrial fibrillation (AF). Many of these patients also have atrial flutter. Therefore, a linear multielectrode has been developed using the same RF energy.The concept and acute results of linear ablation using duty-cycled RF were tested in the cavotricuspid isthmus (CTI).The CTI was targeted in 75 patients, in 68 (90%) among them as an adjunct to AF ablation with the same technology. A linear electrode catheter with a 4-mm tip and five 2-mm ring electrodes was connected to a generator titrating duty-cycled RF at 20-45 W up to a target temperature of 70 degrees C in 1:1 unipolar/bipolar mode.During a mean procedure time of 20 +/- 12 minutes, complete CTI block was achieved by 4 +/- 3 applications of duty-cycled RF in 69 (92%) patients. No more than three RF applications were necessary in 60% of patients. During the initial learning curve, standard RF had to be used in five (7%) patients. Complete block was not achieved in one patient with frequent episodes of AF. Char was observed in five (7%) patients with poor electrode cooling; consequently, the temperature ramp-up was slowed and manually turned off in the event of low-power delivery. Two groin hematomas occurred; otherwise, no clinical complications were observed.Multielectrode catheters delivering duty-cycled RF can effectively ablate the CTI with few RF applications with promising acute results. Further modifications are necessary to improve catheter steering and prevent char formation. (PACE 2010; 444-450).