A Nanometer Resolution Wearable Wireless Medical Device for Non Invasive Intracranial Pressure Monitoring
Open Access
- 18 June 2021
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Sensors Journal
- Vol. 21 (20) , 22270-22284
Abstract
The non-invasive intracranial pressure (NIICP) method based on a skull deformation has been proven to be a significant tool for an assessment of the intracranial pressure (ICP) and compliance. Herein, we present the development and characterization of a novel wireless sensor that uses this method as its working principle and was designed to be easy to use, to have a high resolution, and to achieve a good accessibility. Initially, a brief review of the physiology fundamentals of the ICP and the historic evolution of the NIICP method are mentioned. The sensor architecture and the rationale for the chosen components are then presented, aiming to ensure nanometer displacement measurements, the conversion of analog resolution to digital at a high speed, the fewest amount of distortion, wireless communication, and signal calibration. The NIICP signal has a typical amplitude of $5~\mu \text{m}$ , and thus a resolution of at least 1% of this amplitude is required for an NIICP waveform analysis. We also demonstrate a 40-nm resolution of the sensor using a nanometric displacement test system that can also respond dynamically for NIICP signals from 50 to 180 bpm without any significant distortion (maximum deviation of P2/P1 ratio of 2.6%). The future applications for this device are broad and can enhance a clinical assessment of the intracranial dynamics.Keywords
Funding Information
- brain4care—Braincare Desenvolvimento e Inovação Tecnológica S.A.
- Conselho Nacional de Desenvolvimento Científico e Tecnológico (304312/2020-7i)
This publication has 74 references indexed in Scilit:
- American Society of Neurophysiologic Monitoring and American Society of Neuroimaging Joint Guidelines for Transcranial Doppler Ultrasonic MonitoringJournal of Neuroimaging, 2011
- What Shapes Pulse Amplitude of Intracranial Pressure?Journal of Neurotrauma, 2010
- Brain edema in acute liver failure: Can it be prevented? Can it be treated?Journal of Hepatology, 2007
- Godfrey Hounsfield and the Dawn of Computed TomographyNeurosurgery, 2006
- Assessment of Childhood Intracranial Pressure Recordings Using a New Method of Processing Intracranial Pressure SignalsPediatric Neurosurgery, 2005
- Utility of CSF Pressure Monitoring to Identify Idiopathic Intracranial Hypertension without Papilledema in Patients with Chronic Daily HeadacheCephalalgia, 2004
- Interpretation of ICP Pulse Waves to Determine Intracerebral ComplianceJournal of Neuroscience Nursing, 1988
- Analysis of Pulse Waves of Epidural Pressure Recorded in Acute Intracranial HypertensionNeurologia medico-chirurgica, 1987
- Analysis of Pulse Wave Recorded on Epidural Pressure in Acute Intracranial HypertensionNeurologia medico-chirurgica, 1985
- Analysis of the cerebrospinal fluid pulse wave in intracranial pressureJournal of Neurosurgery, 1983