Measurement of serum albumin by capillary zone electrophoresis, bromocresol green, bromocresol purple, and immunoassay methods
Authors:
- 26 September 2003
- journal article
- research article
- Published by BMJ in Journal of Clinical Pathology
- Vol. 56 (10) , 780-781
Abstract
Background/Aims: The introduction of capillary zone electrophoresis (CZE) to this laboratory has highlighted discrepancies in albumin measured on an Abbott Aeroset by bromocresol green (BCG) and that calculated by CZE on the basis of total protein measured by Biuret. Methods: This study examined differences in albumin estimation by CZE, Abbott Aeroset BCG, and Aeroset bromocresol purple (BCP), and compared these with albumin estimated by Beckman Array immunoassay. Results: Altman and Bland analysis of results showed a positive bias of BCG with CZE (4.51 g/litre; 95% limits of agreement, 3.77 to 5.26; n = 72) and BCP (3.85 g/litre; 95% limits of agreement, −1.42 to 9.12; n = 72). CZE and BCP agreed closely (0.67 g/litre; 95% limits of agreement, −4.39 to 3.06; n = 72). Analysis of 57 of those samples in which BCG and CZE differed ≥ 5 g/litre showed a positive bias of BCG with immunoassay (8.35 g/litre; 95% limits of agreement, 1.54 to 15.16; n =57), with good agreement between CZE and immunoassay (−0.44 g/litre; 95% limits of agreement, −2.82 to 1.94; n = 57). Conclusions: BCP is superior to BCG for the assay of albumin and has replaced BCG as the routine test for albumin in this laboratory.Keywords
This publication has 5 references indexed in Scilit:
- Multicenter evaluation of the Paragon CZETM 2000 capillary zone electrophoresis system for serum protein electrophoresis and monoclonal component typingClinical Chemistry, 1998
- Serum and urine albumin: a progress report on their measurement and clinical significanceClinica Chimica Acta, 1997
- The Measurement of Albumin in Serum and PlasmaAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1985
- Electroimmunoassay of albumin in human serum: accuracy and long-term precision.Clinical Chemistry, 1983
- Albumin by bromcresol green--a case of laboratory conservatism.Clinical Chemistry, 1982